3 operation – Emerson 5081-T User Manual

Page 66

background image

60

9.3

Operation

9.3.1

Off-line and On-line Operation

The Model 375 Communicator features off-line and on-line communications. On-line means the communicator is
connected to the transmitter in the usual fashion. While the communicator is on line, the operator can view meas-
urement data, change program settings, and read diagnostic messages. Off-line means the communicator is not
connected to the transmitter. When the communicator is off line, the operator can still program settings into the
communicator. Later, after the communicator has been connected to a transmitter, the operator can transfer the
programmed settings to the transmitter. Off-line operation permits settings common to several transmitters to be
easily stored in all of them.

9.3.2

Making HART related settings from the keypad

9.3.3

Menu Tree

The menu tree for the Model 375 HART communicator is on the following page. The menu tree for the Model 375
F

OUNDATION

Fieldbus communicator immediately follows.

Note: Model 375 Communicator fully supports F

OUNDATION

Fieldbus on Model 5081-T-FF.

1.

Press MENU. The main menu screen appears. Choose Program.

2. Choose

>>.

3. Choose

HART.

4. To display the device ID, choose DevID. To change the polling address,

choose PollAddrs. To make burst mode settings, choose Burst. To
change the preamble count, choose Preamble.

MODEL 5081-T

SECTION 9.0

OPERATION WITH MODEL 375

Calibrate

Hold

P

Pr

ro

og

gr

ra

am

m

Display

Output

Temp

Measurement

>

>>

>

Security

H

HA

AR

RT

T

>>

D

De

ev

vI

ID

D

PollAddrs

Burst

Preamble