10.6 RTD RESISTANCE VALUES

Table 10-4 is a ready reference of RTD resistance values
at various temperatures. These are used for test and
evaluation of the sensor.

NOTE

Resistance values are read across the RTD
element and are based on the manu-
facturer ’s stated values (±1%). Allow
enough time for the RTD element in the
sensor to stabilize to the surrounding
temperature (10 min).

Temperature Pt-100

Pt-1000

(°C)

Resistance (ohms) Resistance (ohms)

0

100.00

1000

10

103.90

1039

20

107.79

1078

25

109.62

1096

30

111.67

1117

40

115.54

1155

50

119.40

1194

60

123.24

1232

70

127.07

1271

80

130.89

1309

90

134.70

1347

100

138.50

1385

Table 10-4. RTD Resistance Values.

FIGURE 10-5. Conductivity Determination

FORMULA:

EXAMPLE:

cell constant value x 1,000,000

desired simulated conductivity in

µs/cm

.01 x 1,000,000

10

µs/cm

Use the following formula to determine the appropriate resistance value to use to simulate a
conductivity value:

= resistance in ohms

= use 1,000 ohm resistance