beautypg.com

3 temp, U->u9, U7-p – Emerson 5081-T User Manual

Page 59

background image

MODEL 5081-T

SECTION 7.0

PROGRAMMING

7.3 TEMP

7.3.1 Purpose

This section describes how to do the following:

1. Enable and disable automatic temperature compensation

2. Set a manual temperature compensation value for conductivity measurements

3. Tell the transmitter the type of temperature element in the sensor

7.3.2 Definitions

1. AUTOMATIC TEMPERATURE COMPENSATION. Conductivity measurements are directly affected by temperature. A

correction equation in the software automatically corrects for changes caused by temperature. In automatic tempera-
ture correction, the transmitter uses the temperature measured by the sensor for all calculations in which temperature
is used.

2. MANUAL TEMPERATURE COMPENSATION. In manual temperature compensation, the transmitter uses the tem-

perature entered by the user as the reference temperature for corrections of conductivity readings. It does not use the
actual process temperature.

CAUTION

Changing the reference temperature from the default 25°C (77°F) can have large

effects on the conductivity readings and will require different temperature slopes.

7.3.3 Procedure

1.

Press PROG on the remote controller.

2.

Press NEXT until the tEMP submenu appears. Press ENTER.

3.

The screen displays the tAUtO (automatic temperature compensation) prompt.
Press

or

to toggle between On and OFF. Press ENTER to save.

4.

If you disable tAuto, the tMAN prompt appears. Use the arrow keys to change the
temperature to the desired value. To enter a negative number, press

or

until no

digit is flashing. Then press

or

to display the negative sign. The temperature

entered in this step will be used in all measurements, no matter what the
process temperature is.
Press ENTER to save.

5.

Press RESET to return to the process display.

PROGRAM

U->U9

EXIT

ENTER

98

PROGRAM

U7-P

EXIT

ENTER

"

$'c"

PROGRAM

U17:

EXIT

NEXT

ENTER

53