Operation

2

3

4

5

6

7

8

9

10

1

Once you have assembled all the parts and allowed your microscope to come to room temperature, plug the power cord

into the appropriate AC outlet. Note: excess cold can fog lenses, cause lamp to fail, and cause all optics to “fog.” Be sure

to allow time for your scope to acclimate to room temperature.

Turn the light on using the black on/off switch on the rear of the microscope. Next, adjust the light intensity using the

brightness control knob located on the right side of the microscope.

In order to become acquainted with the controls, choose a specimen slide with which you are familiar. For example, an old

hematology slide or a commercially prepared slide. Place the slide into the slide holder by pushing back on the thumb

guard to open the slide finger. The slide finger closes slowly to eliminate the possibility of chipping the corner of your slide

when it closes.

Move the slide to the center of the stage, by turning the stage control knobs, located just below the stage. These knobs

allow you to move the slide on the X-Y axis (forward/backward and left/right).

The sub-stage iris should then be set to match the aperture of the objective for maximum resolution under each objective

power. There are numbers on the iris ring to show the correct setting for each objective power. You should begin with the 4x

or 10x objective. Only use the iris wide open when under the 100x oil objective. Closing down the iris on smaller objective

powers will improve resolution, contrast, and depth of field.

Once you are comfortably seated, look into the oculars and move the eyepiece tubes together or apart until you see only

one complete circle of light. You have now adjusted your interpupillary distance.

Using the 4x or 10x objectives and the coarse and fine adjustment knobs, bring the specimen into focus. Now, move the 40x

objective into place. You will feel a “clicking” action when the objective is seated properly. Again, adjust focus for best

image. You should also adjust the iris diaphragm (as listed above) for the best contrast and resolution.

Diopter Adjustment: Since you are using a binocular or trinocular microscope, you need to adjust for the normal difference in

vision between your two eyes. This is a simple but critical adjustment! First, make sure that the diopter adjustments on the

eyetubes are set to the midpoint of travel, with the “O” mark lined up with the white line near the bottom edge of the

eyetube. Close your left eye and look into the right eyepiece with your right eye. Adjust the fine focus to give you the best

image. Now close your right eye and look with your left eye into the left ocular. Using the diopter adjustment ring on the

ocular tube, adjust the focus until you see a clear, focused field. Now both eyes should see a perfectly focused image.

Friction Adjustment: With repeated use and wear, the stage may drift downward out of focus. If this happens, you need only

to tighten the friction control ring (located on the left side of the microscope between the coarse adjustment and the body

of the microscope). If the coarse focus is hard to turn, you may choose to loosen the friction adjustment.

Parfocality: All LW Scientific microscopes are manufactured to be parfocal – meaning that when you change objectives or

magnification, the specimen will remain very close to being in focus, with only a fine adjustment needed. To achieve the

best parfocality, make certain diopters are set at or near the "0" mark.

Warranty

The Mi5 is covered by a limited lifetime warranty on materials and workmanship, and a 1-year

warranty on electronics from date of purchase. If there is any indication of a problem, contact

LW Scientific. Operating the unit after noticing a problem could compound a simple problem

and cause an unneccessary expense to the owner. LW Scientific support staff will troubleshoot

problems over the phone, and attempt to solve problems in the most expedient manner. This may include sending parts

that can easily be installed by the user, or directing the user through a simple adjustment to the unit. Making repairs to the

unit without authorization from LW Scientific will void the warranty.

If the unit must be shipped in for repair, LW Scientific will issue a Return Merchandise Authorization (RMA) number. You will

need to provide the serial number and have either a warranty card on file at LW Scientific or proof of purchase.