When purchased from LW Scientific or from an authorized distributor, LW Scientific Professional

Stereoscopes and Laboratory Microscopes have a life time warranty against defects in materials

and workmanship and one (1) year warranty on electronic components. All other LW Scientific

instruments, including Student Microscopes and Stereoscopes, have a one (1) year limited

warranty. This warranty is not valid on normal wear and tear, cosmetic damages caused by chemicals, solvents, and/or

cleaning solutions, as well as acts of God.

Please register your product online at: www.lwscientific.com/warranty_form.

Important: Warranty information must be completed within 30 days of purchase.

3

2

5

6

4

1

Always cover your microscope with the dust cover when not in use.

When cleaning the lenses, use lens paper or a cotton swab dipped in lens cleaning solution.

Excess oil should be cleaned off your 100x objective and stage at once. An alcohol pad is best for removing oil from

the stage and on the other metal parts, but is not recommended for use on the lenses. Use lens cleaning solution and

lens paper to clean off your objectives.

Dust in the nosepiece or ocular tubes should be blown out using filtered air. Canned air dusters work well for this job.

Whenever you remove an objective, we recommend that you place it back into the original plastic shipping vial until

ready to be placed back on the microscope. SCREW THE OBJECTIVE SECURELY INTO THE CAP OF THE HOLDER - DO NOT

DROP OBJECTIVE LOOSELY INTO CONTAINER.

To keep your microscope in top condition for years, LW Scientific recommends that you have the microscope

professionally serviced once a year. Warning: The 40x objective is not sealed for oil immersion. Damage to the 40x

objective due to oil immersion is not covered under warranty.

Construction

Rugged alloy; acid and reagent resistant finish

Nosepiece

Quadruple hole multiple ball bearing

Head

Binocular (Seidentopf)

Inclined 30°, rotates 360°

Dual diopter adjustment

(Binocular or trinocular versions)

10X/18 wide-field eyepieces

Monocular or Trinocular

Illumination

Moveable Abbe Condenser (NA 1.25)

Iris Diaphragm

Swing-out filter holder with blue and green filters

20 watt quartz halogen light (12v 20w JC-G4 bulb)

Input 110 and 220V manual switching / 50-60Hz

Objectives

DIN Achromatic,

DIN Semi-Plan or DIN Plan

4X N.A. 0.10

10X N.A. 0.25

40XR N.A. 0.65

100XR N.A. 1.25 (oil immersion)

20X, 50X oil Plan, and 60XR available

Adjustment Controls

Eyepiece: Interpupillary distance adjustment 55-75 mm

Stage Controls: Lateral and forward/reverse positioning;

Etched vernier scales

Coarse Adjustment: Range of 30 mm

Fine Adjustment: Graduation of 2µm

Variable Light Adjustment

Size

N.A.

Mag.

Field of View

4X

0.10

40X

4.5 mm

10X

0.25

100X

1.8 mm

20X

0.40

200X

0.9 mm

40XR

0.65

400X

0.45 mm

50XR

0.95

500X

0.36 mm

60XR

0.85

600X

0.3 mm

100XR

1.25

1000X

0.18 mm

Objectives: The following numbers are

based on use with 10x/18 eyepieces.

Dimensions and Weight

Weight: 14 lbs / 6.4 kg

Height: 15" / 380 mm

Length: 9" / 230 mm

Width: 7" / 178 mm

Shipping weight: 17 lbs/ 8kg

Stage

Mechanical stage

(140 mm x 140 mm) – coaxial drive controls

Range of traverse is 73 mm x 43 mm