Maintenance warranty specifications – LW Scientific Revelation lll Halogen User Manual
Page 4
P. 770.270.1394
F. 770.270.2389
865 Marathon Parkway Lawrenceville GA 30046
Maintenance
Warranty
Specifications
When purchased from LW Scientific or from an authorized distributor, LW Scientific Professional
Stereoscopes and Laboratory Microscopes have a life time warranty against defects in materials
and workmanship and one (1) year warranty on electronic components. All other LW Scientific
instruments, including Student Microscopes and Stereoscopes, have a one (1) year limited
warranty. This warranty is not valid on normal wear and tear, cosmetic damages caused by chemicals, solvents, and/or
cleaning solutions, as well as acts of God.
Please register your product online at: www.lwscientific.com/warranty_form.
Important: Warranty information must be completed within 30 days of purchase.
3
2
5
6
4
1
Always cover your microscope with the dust cover when not in use.
When cleaning the lenses, use lens paper or a cotton swab dipped in lens cleaning solution.
Excess oil should be cleaned off your 100x objective and stage at once. An alcohol pad is best for removing oil from
the stage and on the other metal parts, but is not recommended for use on the lenses. Use lens cleaning solution and
lens paper to clean off your objectives.
Dust in the nosepiece or ocular tubes should be blown out using filtered air. Canned air dusters work well for this job.
Whenever you remove an objective, we recommend that you place it back into the original plastic shipping vial until
ready to be placed back on the microscope. SCREW THE OBJECTIVE SECURELY INTO THE CAP OF THE HOLDER - DO NOT
DROP OBJECTIVE LOOSELY INTO CONTAINER.
To keep your microscope in top condition for years, LW Scientific recommends that you have the microscope
professionally serviced once a year. Warning: The 40x objective is not sealed for oil immersion. Damage to the 40x
objective due to oil immersion is not covered under warranty.
Construction
Rugged alloy; acid and reagent resistant finish
Nosepiece
Quadruple hole multiple ball bearing
Head
Binocular (Seidentopf)
Inclined 30°, rotates 360°
Dual diopter adjustment
(Binocular or trinocular versions)
10X/18 wide-field eyepieces
Monocular or Trinocular
Illumination
Moveable Abbe Condenser (NA 1.25)
Iris Diaphragm
Swing-out filter holder with blue and green filters
20 watt quartz halogen light (12v 20w JC-G4 bulb)
Input 110 and 220V manual switching / 50-60Hz
Objectives
DIN Achromatic,
DIN Semi-Plan or DIN Plan
4X N.A. 0.10
10X N.A. 0.25
40XR N.A. 0.65
100XR N.A. 1.25 (oil immersion)
20X, 50X oil Plan, and 60XR available
Adjustment Controls
Eyepiece: Interpupillary distance adjustment 55-75 mm
Stage Controls: Lateral and forward/reverse positioning;
Etched vernier scales
Coarse Adjustment: Range of 30 mm
Fine Adjustment: Graduation of 2µm
Variable Light Adjustment
Size
N.A.
Mag.
Field of View
4X
0.10
40X
4.5 mm
10X
0.25
100X
1.8 mm
20X
0.40
200X
0.9 mm
40XR
0.65
400X
0.45 mm
50XR
0.95
500X
0.36 mm
60XR
0.85
600X
0.3 mm
100XR
1.25
1000X
0.18 mm
Objectives: The following numbers are
based on use with 10x/18 eyepieces.
Dimensions and Weight
Weight: 14 lbs / 6.4 kg
Height: 15" / 380 mm
Length: 9" / 230 mm
Width: 7" / 178 mm
Shipping weight: 17 lbs/ 8kg
Stage
Mechanical stage
(140 mm x 140 mm) – coaxial drive controls
Range of traverse is 73 mm x 43 mm
MKT-7.5.3-L-055 | Rev 1