Brush Replacement

1

2
3

4

5

6

7

Speed Calibration

With the red high/low switch on the front of the unit on low, the speed dial on maximum, and a tube rotor with tube shields in

place, adjust the VR-I rheostat on the circuit board up to 4000 RPMs.

With the red high/low switch on the front of the unit on high, the speed dial on maximum, and a Hematocrit rotor in place, adjust

the Vr-h rheostat on the circuit board up to 12000 RPMs.

Unplug the centrifuge, and turn it on its side.

Remove the feet and all Phillips head screws holding the metal base plate in place.

Using a flat head screwdriver, remove the two black caps on the bottom part of the motor.

Using your fingers, pull the brush out of the recessed hole that the black caps were removed from. The brushes consist of springs

with a graphite bar on them.

Specifications

Place the new brushes in the recessed hole, and replace the black caps. Be careful not to overtighten. Black caps could break.

(See image).

Replace the metal plate and the feet.

Your Combo V24 is now ready to use.

Max. Speed (HIGH):

Max. Speed (LOW):

Max. Volume:

Input Voltage:

Fuse:

Height:

Width:

Depth:

Weight:

Boxed dimensions:

Boxed weight:

12,000 RPM (+/- 600)

4,000 RPM (+/- 200)

24 3ml to 15ml test tubes

24 75mm capillary tubes

24 2ml microtubes

110v or 220v

8 amp / 250 volt

8.7 inches

10.6 inches

12.5 inches

15.8 lbs.

11.5 x 10 x 7 inches

23 lbs.

MKT-7.5.3-L-076 | Rev 2

RPM’s
500
1,000
1,400
1,750
2,000
2,250
2,500
2,750
3,000
3,250
3,500
3,750
4,000
12,000
12,000

G-Force (RCF)
37
151
296
462
603
764
943
1141
1358
1594
1849
2122
2415
12,900
14,500

Test Tube Rotor

Microtube Rotor
Microhematocrit Rotor

Urine

Blood