Tvoc setup, Selector pins – Ion Science TVOC User Manual

Page 9

background image

TVOC MANUAL

Ion Science Ltd

Page 9 of 32

Unrivalled Detection. www.ionscience.com

Diagram 3

TVOC setup

Selector Pins
TVOC has a number of settings that can be selected by the
user via a row of four selector pins mounted on the reverse side
of the main PCB. Diagram 2 shows the location of the functional
selector pins labelled - A, B, C & D.

The selector pins absence or presence determines the chosen
setting.

TVOC is shipped with all four selector pins fitted as shown in
diagram 3.









Diagram 4 shows the removal of 1 selector pin.








The following table shows the selector pin combinations and the corresponding function.

= Selector pin fitted

 = Selector pin removed


Selector pin

A

B

C

D

Range

Displayed units

1000

ppm (Default)

x

100

ppm

x

10

ppm

x

2280

mg/m

3

x

x

228

mg/m

3

x

x

22.8

mg/m

3

Selector pin ‘B’ and ‘C’ selects the ranges

-

10 ppm, 100 ppm or 1000 ppm

Selector pin ‘D’ selects the units

-

ppm’ or ‘mg/m

3

Selector pin ‘A’

-

Fault output range

NOTE: During an Alarm condition (F1 or F2) the output will drop to either 3.5 mA or 2.0 mA.

Jumper ‘A’ fitted: 3.5 mA during an alarm condition
Jumper ‘A’ removed: 2.0 mA during an alarm condition.
As default, the TVOC has a 100 ppm factory calibration

Diagram 2

Diagram 4