beautypg.com

Ion Science TVOC User Manual

Page 22

background image

TVOC MANUAL

Ion Science Ltd

Page 22 of 32

Unrivalled Detection. www.ionscience.com

Maintenance

5. The PID

sensor’s metal spring

cover can be pulled off using
moderate force. The electrode stack
and PID lamp can then be removed
by holding the PID sensor upside
down.

Warning:
Ensure the electrode stack and
PID Lamp falls onto a soft surface
such as a piece of tissue. This will
avoid damaging the parts as they
fall out and avoids finger contact
with the PID lamp window.


6. Clean the PID Lamp using
Alumina powder loaded onto a
cotton bud. Rub the cotton bud /
Alumina power
on the PID lamp
window
using a circular motion
until a light audible squeak can be
heard.

NOTE:
Contamination of the PID Lamp
window
can considerably reduce
the detection capability of the PID
cell
, even when the contamination
is not visible.

Cleaning of the lamp should be carried out on a regular
basis depending on the duty cycle of the PID lamp
and the environment.




Electrode

stack

PID lamp

Metal spring

cover

Alumina powder

Cotton bud / Alumina

powder

Diagram 17

Diagram 18

Pellet