Technical specifications

PID Sensor

With fence electrode technology

Lamp type

10.6 eV (Krypton)


TVOC enclosure ingress protection

IP65

Sensor ingress protection

IP54


Range

0

– 10 ppm / 0 – 22.8 mg/m

3

(0.01 resolution)

0

– 100 ppm / 0 – 228 mg/m

3

(0.1 resolution)

0- 1,000 ppm / 0

– 2280 mg/m

3

(1.0 resolution)


Non IS applications

Input power

5

– 28 Vdc. 65 mA (0.5mm

2

to 2.5mm

2

C.S.A.)

4

– 20 mA power

8

– 35 Vdc. 22 mA (0.5mm

2

to 2.5mm

2

C.S.A.)

4-20mA loop must be externally powered


ATEX approval

II 2G EEx ia IIC T4 (-

20°C ≤Ta ≤+50°C)

IECEx approval

Ex ia IIC T4 (-20°C Ta +50°C)

Certificate number

Baseefa05ATEX0277

IS applications

Entry parameters

Input power

Ui = 18V

Ii = 800mA

Pi = 1.2W

Ci = 0

F

Li = 0mH

4-20mA power

Ui = 30V

Ii = 200mA

Pi = 1.2W

Ci = 0

F

Li = 0mH

For information only please see ATEX certificate before installation

Zener Barriers:

Required Ask your installation engineers for installation/application advice.



(Non sparking) approval

II 3G Ex nA IIC T4 (-40

o

C < Ta < +50

o

C) (nA for zone 2 only)



Dimensions

Weight

Height

188 mm

(7.40

”)

Instrument

1.3kg

(2.9 lb)

Width

126 mm

(4.96

”)

Packed

1.47kg

(3.2lb)

Depth

78 mm

(3.07

”)


Display

7 Segment, 4 Digit LCD. 4 Colour LED’s

Response

Sensor T90 < 5sec
TVOC output update: Variable 1 second to 10 minutes

Accuracy

0 to 100 ppm: +/- 1 ppm (whichever is greater)

100 to 1000 ppm: +/- 10% or +/- 10 ppm (whichever is greater)

Linearity

0

– 1000 ppm >75%

Calibration

Magnetically accessed

ZERO = Carbon canister
SPAN = 100 ppm Isobutylene +/- 10%

Temperature

Operation: -20

o

C to +50

o

C (-4

o

F to 122

o

F)

Humidity

0

– 95% RH (non-condensing)


EMC

Screened cables are required to archive the light industrial immunity

levels.

NB: All specifications are against an isobutylene calibration at 20 °C, 90% RH and up to 100 ppm.