Technical specifications – Ion Science TVOC User Manual
Page 30
TVOC MANUAL
Ion Science Ltd
Page 30 of 32
Unrivalled Detection. www.ionscience.com
Technical specifications
PID Sensor
With fence electrode technology
Lamp type
10.6 eV (Krypton)
TVOC enclosure ingress protection
IP65
Sensor ingress protection
IP54
Range
0
– 10 ppm / 0 – 22.8 mg/m
3
(0.01 resolution)
0
– 100 ppm / 0 – 228 mg/m
3
(0.1 resolution)
0- 1,000 ppm / 0
– 2280 mg/m
3
(1.0 resolution)
Non IS applications
Input power
5
– 28 Vdc. 65 mA (0.5mm
2
to 2.5mm
2
C.S.A.)
4
– 20 mA power
8
– 35 Vdc. 22 mA (0.5mm
2
to 2.5mm
2
C.S.A.)
4-20mA loop must be externally powered
ATEX approval
II 2G EEx ia IIC T4 (-
20°C ≤Ta ≤+50°C)
IECEx approval
Ex ia IIC T4 (-20°C Ta +50°C)
Certificate number
Baseefa05ATEX0277
IS applications
Entry parameters
Input power
Ui = 18V
Ii = 800mA
Pi = 1.2W
Ci = 0
F
Li = 0mH
4-20mA power
Ui = 30V
Ii = 200mA
Pi = 1.2W
Ci = 0
F
Li = 0mH
For information only please see ATEX certificate before installation
Zener Barriers:
Required Ask your installation engineers for installation/application advice.
(Non sparking) approval
II 3G Ex nA IIC T4 (-40
o
C < Ta < +50
o
C) (nA for zone 2 only)
Dimensions
Weight
Height
188 mm
(7.40
”)
Instrument
1.3kg
(2.9 lb)
Width
126 mm
(4.96
”)
Packed
1.47kg
(3.2lb)
Depth
78 mm
(3.07
”)
Display
7 Segment, 4 Digit LCD. 4 Colour LED’s
Response
Sensor T90 < 5sec
TVOC output update: Variable 1 second to 10 minutes
Accuracy
0 to 100 ppm: +/- 1 ppm (whichever is greater)
100 to 1000 ppm: +/- 10% or +/- 10 ppm (whichever is greater)
Linearity
0
– 1000 ppm >75%
Calibration
Magnetically accessed
ZERO = Carbon canister
SPAN = 100 ppm Isobutylene +/- 10%
Temperature
Operation: -20
o
C to +50
o
C (-4
o
F to 122
o
F)
Humidity
0
– 95% RH (non-condensing)
EMC
Screened cables are required to archive the light industrial immunity
levels.
NB: All specifications are against an isobutylene calibration at 20 °C, 90% RH and up to 100 ppm.