Introduction to tvoc – Ion Science TVOC User Manual

Page 6

background image

TVOC MANUAL

Ion Science Ltd

Page 6 of 32

Unrivalled Detection. www.ionscience.com

Introduction to TVOC

TVOC is a fixed continuous monitor for the detection and measurement of total volatile organic compounds,
which can be dangerous from both a poisoning and explosive perspective. Total VOC’s are detected using
PID (Photo ionisation detection) technology.

TVOC has a number of user selectable features which define its operation and will be application dependent.
The user can define the detection ranges or 0.01 - 10ppm, 0.1

– 100 ppm or 1 – 1,000 ppm (Default);

detection units, either ppm or mg/m

3

; and the duty cycle time i.e. the frequency that the instrument takes a

measurement. Duty cycles are explained in more detail under TVOC set up, and Operation.

IMPORTANT: TVOC can be calibrated on alternative gases to Isobutylene however to ensure TVOC
remains within specification, the correct concentration for the alternative gas must be chosen.
Example: Benzene gas has a response factor of 0.5 which means only half as much concentration is
required to produce the equivalent to Isobutylene.

The ideal gas / concentrations used to set the calibration SPAN are shown below:-

Lower limit

Upper limit

10 ppm range

8 ppm

12 ppm

Isobutylene equivalent

100 ppm range

90 ppm

110 ppm

Isobutylene equivalent

1,000 ppm range

90 ppm

110 ppm

Isobutylene equivalent



How to calculate alternative concentration of gases for gases other than Isobutylene:-
Calibration

Ideal response

Response factor

Ideal calibration gas

Gas

for Isobutylene

concentration

Benzene

100

x

0.5

=

50 ppm

Isoprene

100

x

0.7

=

70 ppm

Isobutylene

100

x

1.0

=

100 ppm

Carbon disulfide

100

x

1.4

=

140 ppm

Ethyl acrylate

100

x

2.0

=

200 ppm


For other gas response factors please refer to the Ion Science web site or contact Ion Science Ltd.
http://www.ionscience.com/GasSearch/tabid/87/Default.aspx

WARNING: Ignoring these guide lines may result in calibration failure, loss of resolution or loss of range.

The default settings of TVOC are:

Detection range

1

– 1000 ppm

Units

-

ppm

Signal update

-

60 seconds


TVOC gives a continuous 4-20 mA output which can be integrated into a DCS control system to give an
indication of VOC levels in the operating environment. Note the 4-20 mA output must be externally powered
with 8-35 V.

In addition to the 4-20 mA output TVOC has an LCD display showing gas concentration and 4 c

olour LED’s.

LED functions are as follows:

Green

-

is an indicator of TVOC’s working status

Red

- is the fault indicator

Yellow x 2

- are calibration status indicators


Please see Calibration for more information on LED status.