Calibration

Calibration routine


Ion Science recommends calibrating TVOC after any maintenance or lamp cleaning is carried out and on a 3
monthly basis to ensure TVOC is working to specification.

NOTE: Please read this entire calibration procedure before attempting a calibration.

TVOC calibration is undertaken in four steps:

1. Setting the ZERO

(Using clean gas via carbon filter)

2. Setting the gas tolerance

(1

– 200 ppm Isobutylene on 0-100 ppm and 0-1000 ppm ranges)

(0.1

– 20.0 ppm Isobutylene on the 0-10ppm range)

3. Setting the SPAN

(Using the SPAN gas)
















Setting the ZERO

1

Place the calibration magnet over the top right hand corner of the Ion Science logo to enter
calibration mode. The

ZERO LED’ will illuminate and the ‘STATUS LED’ will extinguish.

2

Remove the calibration magnet

3

Insert the calibration adaptor into the sensor cap, attach the carbon filter to the calibration adaptor.
The carbon filter should be attached to the hand aspirator.

4

Again place the calibration magnet over the logo. The

ZERO LED’ will flash, during this stage

TVOC displays a direct millivolt (mV) output from the PID sensor.









STATUS LED


ZERO LED




SPAN LED


FAULT LED

Calibration adaptor

Carbon filter

Hand
aspirator

Diagram 11

D

iagram 12: LED’s

STATUS, ZERO, SPAN, FAULT

Diagram 13