Calibration routine, Setting the zero – Ion Science TVOC User Manual
Page 18
TVOC MANUAL
Ion Science Ltd
Page 18 of 32
Unrivalled Detection. www.ionscience.com
Calibration
Calibration routine
Ion Science recommends calibrating TVOC after any maintenance or lamp cleaning is carried out and on a 3
monthly basis to ensure TVOC is working to specification.
NOTE: Please read this entire calibration procedure before attempting a calibration.
TVOC calibration is undertaken in four steps:
1. Setting the ZERO
(Using clean gas via carbon filter)
2. Setting the gas tolerance
(1
– 200 ppm Isobutylene on 0-100 ppm and 0-1000 ppm ranges)
(0.1
– 20.0 ppm Isobutylene on the 0-10ppm range)
3. Setting the SPAN
(Using the SPAN gas)
Setting the ZERO
1
–
Place the calibration magnet over the top right hand corner of the Ion Science logo to enter
calibration mode. The
‘ZERO LED’ will illuminate and the ‘STATUS LED’ will extinguish.
2
–
Remove the calibration magnet
3
–
Insert the calibration adaptor into the sensor cap, attach the carbon filter to the calibration adaptor.
The carbon filter should be attached to the hand aspirator.
4
–
Again place the calibration magnet over the logo. The
‘ZERO LED’ will flash, during this stage
TVOC displays a direct millivolt (mV) output from the PID sensor.
STATUS LED
ZERO LED
SPAN LED
FAULT LED
Calibration adaptor
Carbon filter
Hand
aspirator
Diagram 11
D
iagram 12: LED’s
STATUS, ZERO, SPAN, FAULT
Diagram 13