Measured value output – HEIDENHAIN ND 920 User Manual
Page 59

59
Delay times for data output
Storage delay
t
1
≤
0.5 ms
Data output after
t
2
≤
30 ms
Regeneration time
t
3
≥
0 ms
Duration of data output in seconds:
t
D
=
Next possible signal for measured value output:
t
E
= t
1
+t
2
+ t
D
+ t
3
[s]
Measured value output with CTRL B
If the control character STX (CTRL B) is received over the RS-232-C
interface, the measured value referenced to this time point will be sent
over the interface. CTRL B is received over RXD and the measured
values are output over TXD.
t2
t3
tD
CTRLB
CTRLB
t1
176 x number of axes + 11 x number of blank lines
Baud rate
BASIC program for measured value output:
10
L%=48
20
CLS
30
PRINT "V.24/RS232"
40
OPEN "COM1:9600,E,7" AS#1
50
PRINT #1, CHR$ (2);
60
IF INKEY$<>""THEN 130
70
C%=LOC(1)
80
IF C% 90 X$=INPUT$(L%,#1) 100 LOCATE 9,1 110 PRINT X$; 120 GOTO 50 130 END Measured Value Output