Operating Parameters
P60.0 to P60.7
Activate switching range for EXT
connection and assign to the axes
2)
No switching range
SWITCH OUT.0 OFF
Switching range for X axis
SWITCH OUT.0 X
Switching range for Y axis
SWITCH OUT.0 Y
Switching range for Z axis
SWITCH OUT.0 Z
P61.0 to P61.7
Define switching range for EXT
connection
2)
Enter switching point (= display value): the switching range is
symmetrical about the display value 0.
Input range
(
in millimeters
)
:
0 to 99 999.999
P81.1 to P81.3
Encoder
Max. encoder signal 16 µA
pp
ENCODER X 16µA
Max. encoder signal 40 µA
pp
ENCODER X 40µA
P96 Measured value output with probing
2)
Measured value output active
PROBE RS232 ON
Not active
PROBE RS232 OFF
P97 Code for measured values
2)
ASCII character for identifying measured values for measured
value output with probing, contact or pulse
Input range:
0 to 127
No ASCII character
0
ASCII character from ASCII table
1 to 127
P98 Dialog language
1)
German
DIALOG LANG. D
English
DIALOG LANG. GB
French
DIALOG LANG. F
Italian
DIALOG LANG. I
Dutch
DIALOG LANG. NL
Spanish
DIALOG LANG. E
Danish
DIALOG LANG. DK
Swedish
DIALOG LANG. S
Czech
DIALOG LANG. CZ
Japanese
DIALOG LANG. J
1)
User parameter
2)
Only on units with RS-232-C/V.24 and EXT connection