Switching Inputs and Outputs EXT (X41)

Inputs

Input signals

Internal pull-up resistor 1 k

W, active with low level

Trigger by making contact against 0 V or
by low level signal over TTL logic device

Minimum pulse duration: t

³ 30 ms,

with fast set/zero reset: t

³ 30 µs

Minimum pulse duration: t

³ 30 ms, with set/zero reset:

t

³ 1.5 ms, with fast set/zero reset: t ³ 30 ms

Delay for set/zero reset: fast data output t

d

£ 25 µs;

slow data output t

d

£ 2 ms

B

UCE

E

C

Pin 1.10

0 V

+ UB ≤ 32 V

I

≤ 100 mA

t

min

³ÿ30 ms

0V

E

0V

E

t

min

Outputs

Output signals

Open collector outputs, active with low level

Delay until signal output: t

d

£ 30 ms

Signal duration of zero signal, trigger limit A1, A2: t

0

³ 180 ms

Signal level of outputs

Status

Level

High

U

£ + 32 V; I £ 10 µA

Low

U

£ + 0.4 V; I £ 100 mA

Signal level of inputs

Status

Level

High

+ 3.9 V

£ U £ + 15 V

Low

– 0.5 V

£ U £ + 0.9 V; I £ 6 mA