List of operating parameters – HEIDENHAIN ND 282 B User Manual
Page 22

22
Parameter
Settings / Function
P57 BCD
Behavior without latch signal
Data output always active
TRIST. ON
High-impedance output (3-state) TRIST. OFF
P62 A1
Trigger limit 1
P63 A2
Trigger limit 2
P79 PRESET
Value for datum point
Enter numerical value for
datum setting over switching input
or with ENT key
P80 ENT-CL
Set display
No set/Set with
CL/ENT
CL-ENT OFF
Zero reset with CL
setting disabled with ENT
CL......ON
Zero reset with CL and set with
ENT to value selected in P79
CL-ENT ON
P82 DISPL.ON
Message after switch-on
ENT...CL message
MESSAGE ON
No message
MESSG. OFF
P84 BCD
Display errors
Error display on
ERROR ON
Error display off
ERROR OFF
Parameter
Settings / Function
P44 REF
Reference mark evaluation
Evaluate the reference mark
REF. ON
Do not evaluate the reference
mark
REF. OFF
P45 ALARM
Encoder monitoring
No monitoring
ALARM OFF
Monitor the frequency
FREQUENCY
Monitor contamination
CONTAMINAT.
Contamination + frequency FRQ. + CONT.
P53 BCD
Data output speed
Fast
BCD FAST
Slow
BCD SLOW
P54 BCD
Latch rate
LATCH 0.2 / 0.4 / 0.8 / 1.6 /
3.2 / 6.4 / 12.8 / 25.6
[µs]
P55 BCD
Data output during value display
Concurrent
BCD ACTL.
Frozen / hold
BCD HOLD
Frozen / concurrent
BCD STOP
P56 BCD
Algebraic sign level
Low = Minus
MINUS LOW
High = Minus
MINUS HIGH
List of Operating Parameters