BCD Data Interface (X33)

There are several ways to start measured value output:

ä Press MOD repeatedly until the PRINT indicator blinks

(only for“slow” data output), then start data output with
ENT;
or

ä Let the measured values periodically be sent to the data

output;
or

ä nput a latch command at the D-sub connection EXT;

or

ä nput a latch command at the BCD connection.

Interface Operating Mode (see Operating Parameter P53)
slow–

Display value

fast –

Display value referenced to datum 1
(MIN/MAX/DIFF display values are not shown)

You can order a connecting cable (to a PC, for example) from
HEIDENHAIN (Id. Nr. 206 420-..); maximum cable length 10 m.

Operating parameters for display output: P23, P53 to P57,
P84.

BCD Data Interface (X33)

“AMP-Champ“ Connector (36-pin, female)

Pins

Assignment

2

0

2

1

2

2

2

3

1

2

3

4

Decade 1

5

6

7

8

Decade 2

9

10

11

12

Decade 3

13

14

15

16

Decade 4

17

18

19

20

Decade 5

21

22

23

24

Decade 6

25

26

27

28

Decade 7

29

30

31

32

Decade 8

Pins

Assignment

33

Algebraic sign

34

Ready message

35

Display value

36

0V

Output Level

Low: U £ 0,4 V at I £ 6 mA

High: U ³ 3,8 V at I £ 2,6 mA
The output signals are TTL
compatible.

Latch level

Low: U £ 0,9 V at Imax £ 6 mA

High: U ³ 3,9 V; or
TTL level (internald pull-up
resistor 10 k

W).