Parameter
Settings / Function
P00 CODE
Enter a code number:
9 51 48:
To change the protected
operating parameters
24 65 84: To lock the keyboard
66 55 44: To show the software version
24 65 82: Distance-to-go display
P01
Units of measurement
Display in millimeters
MM
Display in inches
INCH
P17 CLASS.
Sorting and tolerance checking
Sorting and tol. checking ON
CLASS. ON
Sorting and tol. checking OFF CLASS. OFF
P18 L.CLASS.
Lower limit for sorting
P19 U.CLASS.
Upper limit for sorting
P21 M.SERIE
Display for series of measurements
OFF
DIFF MAX ACTL MIN
P23 DISPL.
Display stop for measured value output
Display equal to actual valueDISPL. ACTL.
Display equal to output
DISPL. BCD
List of Operating Parameters
List of Operating Parameters
Parameter
Settings / Function
P30 DIR
Counting direction
Positive counting direction with
positive direction of traverse DIRECT. POS
Negative counting direction with
positive direction of traverse DIRECT. NEG
P32 SUBD.
Encoder signal subdivisions
200 / 100 / 50 / 40 / 20 / 10 / 8 / 5 / 4 / 2 /
1 / 0.8 / 0.5 / 0.4 / 0.2 / 0.1
P33 COUNT
Counting mode
0-1-2-3-4-5-6-7-8-9
COUNT 0-1
0-2-4-6-8
COUNT 0-2
0-5
COUNT 0-5
P38 DEC.
Decimal places
1 / 2 / 3 / 4 / 5 / 6
(up to 7 for inch display)
P43 REF
Reference marks
One reference mark
SINGLE REF.
Distance-coded with 500 • SP
(SP: signal period)
500 SP
Distance-coded with 1000 • SP
(e.g. for HEIDENHAIN LS ...C)
1000 SP
Distance-coded with 2000 • SP
2000 SP
Distance-coded with 5000 • SP
5000 SP