Error messages – HEIDENHAIN ND 282 B User Manual
Page 15

15
Error Messages
Error Messages
Display
Display
Effect/Cause
BCD Output
2)
BCD SPEED
Last measured value not yet
01
called (with external data
requesting).
1)
NO RES.PRES
Illegal attempt to set or zero reset.
04
The display will not be set
or zero reset.
1)
OVERFLOW
Input value not displayable
12
(input through keys or external
setting).
1)
Internal counteroverflow
53
(BCD counter).
1)
SIGNAL
Encoder signal is too weak.
50
The scale may be contaminated.
1)
FREQUENCY
Input frequency too high for encoder 51
input. Traversing speed may be
too fast.
1)
ERR. REF.
The spacing of the reference marks 55
as defined in P43 is not the actual
spacing.
1)
Display
Display
Effect/Cause
BCD Output
2)
ERR. MEMORY
Checksum error: Check the datum,
80
operating parameters and compen-
sation values for non-linear axis error
compensation. If this error recurs,
contact your service agency!
1)
These errors are important for the attached device. The
error signal (pin 19) at the D-sub connection EXT is active.
2)
Active with P84: ERROR ON
The ND sends the error message to the BCD output on
the first and second decades. All other decades receive the
ASCII character “A“ (1010).
Other Error Displays
If “OVERFLOW” appears, the measured value is too large or
too small:
ä Set a new datum.
or
ä Traverse back.
If
all sorting signals light up, the upper sorting limit is
smaller than the lower limit:
ä Change operating parameters P18 and/or P19.
To clear error messages:
Once you have removed the cause of the error:
ä Clear the error message with the CL key.