beautypg.com

Error messages – HEIDENHAIN ND 282 B User Manual

Page 15

background image

15

Error Messages

Error Messages

Display

Display

Effect/Cause

BCD Output

2)

BCD SPEED

Last measured value not yet

01

called (with external data

requesting).

1)

NO RES.PRES

Illegal attempt to set or zero reset.

04

The display will not be set

or zero reset.

1)

OVERFLOW

Input value not displayable

12

(input through keys or external

setting).

1)

Internal counteroverflow

53

(BCD counter).

1)

SIGNAL

Encoder signal is too weak.

50

The scale may be contaminated.

1)

FREQUENCY

Input frequency too high for encoder 51

input. Traversing speed may be

too fast.

1)

ERR. REF.

The spacing of the reference marks 55

as defined in P43 is not the actual

spacing.

1)

Display

Display

Effect/Cause

BCD Output

2)

ERR. MEMORY

Checksum error: Check the datum,

80

operating parameters and compen-

sation values for non-linear axis error

compensation. If this error recurs,

contact your service agency!

1)

These errors are important for the attached device. The
error signal (pin 19) at the D-sub connection EXT is active.

2)

Active with P84: ERROR ON

The ND sends the error message to the BCD output on
the first and second decades. All other decades receive the
ASCII character “A“ (1010).

Other Error Displays
If “OVERFLOW” appears, the measured value is too large or
too small:
ä Set a new datum.

or

ä Traverse back.

If

all sorting signals light up, the upper sorting limit is

smaller than the lower limit:
ä Change operating parameters P18 and/or P19.

To clear error messages:

Once you have removed the cause of the error:
ä Clear the error message with the CL key.