Instruction manual – Emerson NGA 2000 TO2 User Manual
Page 65
Instruction Manual
760008-A
September 2001
Rosemount Analytical Inc. A Division of Emerson Process Management
Maintenance and Service 4-5
Model NGA 2000 TO2
Select Linearization…
Select Linearization raw values (sensor cur-
rent)…
Enter the data for each raw value point.
Return to the previous menu (Back…)
Select Linearization set point values (ppm
values)…
Enter the data for each set point value.
Return to the previous menu (Back…)
Select Linearization temperature values…
Enter the data for each temperature value
point..
Return the the previous menu (Back…)
Select Load linearization data! press enter.
The new data is now loaded into the analyzer
and accurate measurements can be obtained.
TO2
37.50 ppm
-- Analyzer Module Setup and Controls --
Calibration…
Alarms…
Ranges…
Linearization…
Measurement display configuration…
Concentration measurement…
Flow measurement…
Temperature measurement…
Measure
Back…
Status…
TO2
37.50 ppm
-- View & Modify Linearization Table --
Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…
Load linearization data!
Measure
Back…
TO2
37.50 ppm
-- Linearization Raw Values --
X1: 0.3828 uA
X2: 3.0400 uA
X3: 20.9300 uA
X4: 193.000 uA
Measure
Back…
TO2
37.50 ppm
-- View & Modify Linearization Table --
Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…
Load linearization data!
Measure
Back…
TO2
37.50 ppm
-- Linearization Set Point Values --
Y1: 0.0000 ppm
Y2: 1.050 ppm
Y3: 7.000 ppm
Y4: 72.800 ppm
Measure
Back…
TO2
37.50 ppm
-- View & Modify Linearization Table --
Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…
Load linearization data!
Measure
Back…
TO2
37.50 ppm
-- Linearization Temperature Values --
T1: 25.0 °C
T2: 25.0 °C
T3: 25.0 °C
T4: 25.0 °C
Measure
Back…
TO2
37.50 ppm
-- View & Modify Linearization Table --
Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…
Load linearization data!
Measure
Back…