beautypg.com

B. single component display, 50 % co, Instruction manual – Emerson NGA 2000 TO2 User Manual

Page 33

background image

Instruction Manual

760008-A

September 2001

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-9

Model NGA 2000 TO2

b. Single Component Display

The Measure menu that displays after
startup is the Single Component display
of the analyzer. If other analyzer modules

are connected to the Platform, it is possi-
ble to display them using the following
steps to change the channel of the single
component display:

Figure 3-11. Single Component Display

Press F4 (Channel) to change to the Single
Component Display of any other installed
Analyzer Modules.

Example:
Changing from TO2 (channel 1) to CO

2

(channel 2).

Note: Display may look different depending
on installed analyzers.

Continue pressing F4 to display the desired
channel depending on the installed analyzer
configuration, ultimately returning to the first
channel.

Single component display of the starting
channel.

TO2

Display

Status…

Main…

Channel

BasicCal

0.00 Range: 3 50.00

37.50

ppm O2

Electrolyte level: OK

Temp current: 263 uA

0.0

600

Range: 3

Sensor Current: 0.0500 uA 0.0000 100.0000

MLT/CH1/R2

Display

Status…

Main…

Channel

BasicCal

0.00 Range: 2 5.00

2.50 % CO

2

Temperature: 37.0 C

0.0

Maintenance-Requests: No

Any-alarms: No

Operation: Ready

TO2

Display

Status…

Main…

Channel

BasicCal

0.00 Range: 3 50.00

37.50

ppm O2

Electrolyte level: OK

Temp current: 263 uA

0

600

Range: 3

Sensor Current: 0.0500 uA 0.0000 100.0000