B. analog output setup, Select analog output number, Choose analyzer module – Emerson NGA 2000 TO2 User Manual

Page 50: Choose signal, Instruction manual

background image

Instruction Manual

760008-A
September 2001

3-26 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 TO2

b. Analog Output Setup

In the System SIO Module menu (Figure
3-25), select Analog output setup…

Select analog output number

Choose the desired analog output (1-8) to
set the parameters. The number of out-
puts depends on the analyzer configura-
tion as 2, 4, 6, or 8.

Choose the analyzer module

Select the “Analyzer Modules” submenu
by selecting the “Choose signal source
module…” line and pressing the

↵ key.

Select the tag of the desired reference
channel with the

↑ or ↓ keys and then

press the

↵ or → key. The display will re-

turn to the previous menu automatically
and the selected reference channel will be
displayed in the “Source module:” line.

Choose signal

Select the “Signals” submenu by select-
ing the “Choose signal…” line and press-
ing the

↵ key.

Note: the parameters “Operation status”
and “Validity” are not functional on the
TO2. The list of signals will depend on the
module chosen.

Pr

Press the F5 key to go to a further menu
to choose the Primary Variable signal for
the analog output. The Primary Variable is
the actual oxygen concentration.

The signal chosen here will be applied to
the analog output (1-8) chosen above.

TO2

37.50 ppm

-- Analog Output Setup --

Output number: 1

Choose signal source module…
Choose signal…
Signal value for 0% output: 0.00
Signal value for 100% output: 100.00
Output current: 0…20 mA
Hold output during calibration: No
Signal name: Sample flow
Current signal value: 8.60
Source module: TO2

Measure

Back…

More…

TO2

37.50 ppm

-- Analog Output Setup --

Output number: 1

Choose signal source module…

Choose signal…
Signal value for 0% output: 0.00
Signal value for 100% output: 100.00
Output current: 0…20 mA
Hold output during calibration: No
Signal name: Sample flow
Current signal value: 8.60
Source module: TO2

Measure

Back…

More…

TO2

37.50 ppm

Measure

Back…

-- Analyzer Modules --

TO2 1.0

MLT/CH1
MLT/CH2
MLT/CH3

<<<

>>>

TO2

37.50 ppm

Measure

Back…

-- Signals --

Sensor temp

Temp current

Sensor current

Sample flow

Range

Electrolyte level

Operation status

Validity

<<<

>>>

TO2

37.50 ppm

Measure

Back…

-- Signals --

Primary variable

<<<

>>>