Output(s) value on analyzer failure, Output number, Operation mode – Emerson NGA 2000 TO2 User Manual

Page 52: Instruction manual

background image

Instruction Manual

760008-A
September 2001

3-28 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 TO2

Output(s) value on analyzer
failure

Choose the desired signal level to cause
a failure condition. The choices are:

Actual

BeginOfRange

EndOfRange

BeginOfRange-10%

BeginOfRange+10%

Output number

Choose the output number (1-8) for set-
ting the fine adjustment.

Operation mode

Normal: The absolute measurement sig-
nal will be sent to the analog output.

Adjust 0V: Used to set the display equal
to the analog output for 0V and 0 mA. Life
zero signals (4-20 mA and 2-10V) are set
automatically and cannot be adjusted.

Adjust 10V: Used to set the display equal
to the analog output for 10V and 20 mA.

Select the “Fine adjustment for 0% out-
put” and/or “Fine adjustment for 100%
output” lines with the

↵ or → key. Adjust

to the desired value with the

↑ or ↓ key

and confirm with the

↵ key. The range of

values are:

3000 to 6000 for 0% (default 4096)

600 to 1000 for 100% (default 819)

The last three lines of the “Analog Output
Setup” menu are display only for configu-
ration values of the analog output.

Signal name: The name of the sig-
nal chosen in the “Choose signal”
menu.

Current signal value: The current
value of the variable.

Source module: The name of the
module chosen in the “Choose sig-
nal source module” menu.Pressing
the F5 (More…) key changes to the
submenu “Special Scaling for Con-
centration Signal.”

This menu allows for the setting of each
of the 8 outputs to be the same as the
range limits “Yes” or as set on the previ-
ous menus.

See section 3-5c for setting the range lim-
its.

Pressing F5 (More…) changes to the
submenu “Special Scaling for Concentra-
tion Signal”.

TO2

37.50 ppm

Measure

Back…

-- Output Signal if Assigned Module Fails --

Output(s) value on analyzer failure: BeginOfRange – 10%

-- Fine Adjustment –

Output number: 1
Operation mode: Normal
Fine adjustment for 0% output: 4096
Fine adjustment for 100% output: 819

More…

TO2

37.50 ppm

Measure

Back…

-- Output Signal if Assigned Module Fails --

Output(s) value on analyzer failure: BeginOfRange – 10%

-- Fine Adjustment –

Output number: 1

Operation mode: Normal
Fine adjustment for 0% output: 4096
Fine adjustment for 100% output: 819

More…

TO2

37.50 ppm

Measure

Back…

-- Output Signal if Assigned Module Fails --

Output(s) value on analyzer failure: BeginOfRange – 10%

-- Fine Adjustment –

Output number: 1

Operation mode: Normal

Fine adjustment for 0% output: 4096
Fine adjustment for 100% output: 819

More…