SL Power Electronics MINT1065B Series User Manual
Page 2

Attention, Consult
Accompanying Documents
Alternating Current
Attention, Dangerous Voltages
EXPLANATION OF SYMBOLS
Direct Current
Class II
OUTPUTS:
The output must not be connected to Earth ground or to metallic enclosure in the end application.
The output has not been evaluated for direct patient connection (consult factory for details). The DC output is
SELV under normal and single fault conditions.
OVERVOLTAGE PROTECTION:
The output is monitored for an overvoltage condition. In some
applications where an overvoltage condition could result in a hazard as defined in applicable safety standards,
redundant or additional overvoltage protection may be required. Consult factory for details.
CAUTION: When performing Dielectric Strength Tests, catastrophic failure of the unit may result if a
Dielectric Strength test voltage greater than 1800 V ac is applied between primary and secondary circuits. The
components providing isolation from primary to secondary cannot be tested while installed in the power supply
without overstressing basic (primary to ground) insulation. All isolating components are individually 100 %
tested at 4800 V ac prior to installation.
ISOLATION:
The creepage distance between primary and secondary circuits is 8 mm minimum. The
required creepage and clearance distances from primary circuits to secondary circuits must be maintained after
installation to preserve the intended safety. This may require the use of plastic attaching hardware and standoffs
in the end application. The use of an insulator between the PWB Bottom and a metal chassis should be
considered.
TEMPERATURES: The maximum operating temperatures of certain safety components, as defined in the
applicable safety standards, must not be exceeded after installation to preserve the intended safety. The output
power, ambient air temperature and the availability, amount, direction and/or restriction of airflow influence the
temperatures of these components.
OVERCURRENT PROTECTION: The MINT1065B series has dual internal fusing as required by IEC/EN/UL
60601-1.
WARNING! RISK OF FIRE! A blown internal fuse is an indication of catastrophic failure of circuit
component(s). Refer to fuse marking on the supply for rating. Repair must be performed by SLPE authorized
personnel.
WARNING! SHOCK HAZARD! Dangerous voltages are present on some components, printed wiring traces
and heatsinks.
C O N N E C T I O N S
SLPE will not be liable for the safety, reliability or performance of these power supplies if a) any changes, modifications or repairs are carried out by other
than authorized agents of SLPE, or b) the installation of the supply is not in accordance with these installation instructions and the applicable UL, CSA, and
EN/IEC safety standards
.
41-36127-0002-1
Page 2 of 2
J1 Pin
AC Input
J2 Pin
DC Output
1
Line
1
Output #1 (+)
3
Neutral
2
Output #1 (+)
3
Return
4
Return
RECOMMENDED MATING CONNECTORS
J1 AMP 640250-3 (center pin not used)
J2 AMP 640250-4
CAUTION
Do not exceed 5 Amps per contact