SL Power Electronics G2T60 User Manual
G2t60 series installation instructions
![background image](https://www.manualsdir.com/files/804872/content/doc001.png)
SL POWER ELECTRONICS CORP.
6050 King Drive, Bldg. A
Ventura, CA 93003 + 805-486-4565
Internet: www.slpower.com
MODEL NUMBERS: G2T60-X-YYYG, where X = 12, 15, 18, 24 or 48 which represents the output
voltage rating –YYY is value added options not related to Safety and G represents RoHS Compliance.
RATINGS:
Input:
100-240 V ac, 1.5 A, 50/60 Hz
Output:
Notes:
1.
Maximum Operating Relative Humidity 96 %, no condensation.
2. Storage: -40 to +85 °C. Units should be allowed to warm-up under non-condensing conditions before application of
power.
CERTIFICATION: All models are Certified to be in compliance with the applicable requirements of
EN/IEC/UL 60601-1 and CSA 22.2 No. 601.1. In addition, all models (except G2T60-18) are Certified
to be in compliance with EN/IEC/UL/CSA 60950-1.
CLASSIFICATION:
(5.1) Protection against electric shock = Class II
(In accordance with sub-
(5.2) Degree of protection against electric shock = Signal output or intermediate
clause 5 of IEC 60601-1)
(5.3) Protection against harmful ingress of water = Ordinary (no protection)
(5.5) Have not been evaluated for use in the presence of a flammable anaesthetic
mixture with air, oxygen, or nitrous oxide. This evaluation is to be made
on the end equipment by the OEM.
(5.6) Mode of operation = Continuous
SAFETY DECLARATION: SL Power Electronics Corp. (SLPE) declares under our sole
responsibility that all models listed above are in conformity with the applicable
requirements of EN 60950-1 following the provisions of the Low Voltage Directive
2006/95/EC.
OUTPUT: The output must not be connected to Earth ground or to metallic enclosure in the end
application. The output has not been evaluated for direct patient connection (consult factory for details).
The DC output is SELV under normal and single fault conditions.
OVERVOLTAGE PROTECTION: The output is monitored for an overvoltage condition. In some
applications where an overvoltage condition could result in a hazard as defined in applicable safety
standards, redundant or additional overvoltage protection may be required. Consult factory for details.
41-35510-0002 Rev. D 08/18/2008
Page 1 of 2
G2T60 SERIES
INSTALLATION INSTRUCTIONS
MODEL
40 °C Ambient
Convection Cooled
50 °C Ambient
Convection Cooled
50 °C Ambient
100 LFM Airflow
G2T60-12
12 V dc 4.2 A
12 V dc 3.35 A
12 V dc 5.0 A
G2T60-15
15 V dc 4.0 A
15 V dc 3.7 A
15 V dc 4.35 A
G2T60-18
18 V dc 3.35 A
18 V dc 3.1 A
18 V dc 3.6 A
G2T60-24
24 V dc 2.1 A
24 V dc 1.9 A
24 V dc 2.7 A
G2T60-48
48 V dc 1.35 A
48 V dc 1.25 A
48 V dc 1.35 A