SL Power Electronics GNT60 User Manual

Gnt60 series installation instructions

background image

SL POWER ELECTRONICS CORP.
6050 King Drive, Bldg. A
Ventura, CA 93003 + 805-486-4565
Internet: www.slpower.com

MODEL NUMBERS: GNT60-X-YYYG, where X = 12, 15, 18, 24 or 48 which represents the output
voltage rating, -YYY is value added options not related to Safety and G represents ROHS Compliance


RATINGS:

Input:

100-240 V ac, 1.5 A, 50/60 Hz

Output:








Notes:
1. Maximum Operating Relative Humidity 96 %, no condensation.
2. Storage: -40 to +85 °C. Units should be allowed to warm-up under non-condensing conditions before application of

power.

CERTIFICATION: All models are Certified to be in compliance with the applicable requirements of
CSA 22.2 No. 601.1 and EN/IEC/UL 60601-1. In addition, all models (except GNT60-18) are Certified
to be in compliance with EN/IEC/UL/CSA 60950-1.

CLASSIFICATION:

(5.1) Protection against electric shock = Class I

(In accordance with sub-

(5.2) Degree of protection against electric shock = Signal output or intermediate

clause 5 of IEC 60601-1)

(5.3) Protection against harmful ingress of water = Ordinary (no protection)

(5.5) Have not been evaluated for use in the presence of a flammable anaesthetic

mixture with air, oxygen, or nitrous oxide. This evaluation is to be made
on the end equipment by the OEM.

(5.6) Mode of operation = Continuous

SAFETY DECLARATION: SL Power Electronics Corp. (SLPE) declares under our sole
responsibility that all models listed above are in conformity with the applicable
requirements of EN 60950-1 following the provisions of the Low Voltage Directive

2006/95/EC.

GROUNDING:

The ground connection (quick connect tab) must be bonded to protective earth in the

end equipment. Using this terminal for the system ground is not recommended. A separate dedicated
grounding point should be used.

OUTPUT: The output is intended for Protectively Earthed Signal Output and Intermediate Circuits
only. The output is not acceptable for patient connection without additional isolation. The DC output is
SELV under normal and single fault conditions.

OVERVOLTAGE PROTECTION: The output is monitored for an overvoltage condition. In some
applications where an overvoltage condition could result in a hazard as defined in applicable safety
standards, redundant or additional overvoltage protection may be required. Consult factory for details.

41-35510-0001 Rev. D 08/15/2008

Page 1 of 2

GNT60 SERIES

INSTALLATION INSTRUCTIONS

MODEL

40 °C Ambient

Convection Cooled

50 °C Ambient

Convection Cooled

50 °C Ambient

100 LFM Airflow

GNT60-12

12 V dc 4.2 A

12 V dc 3.35 A

12 V dc 5.0 A

GNT60-15

15 V dc 4.0 A

15 V dc 3.7 A

15 V dc 4.35 A

GNT60-18

18 V dc 3.35 A

18 V dc 3.1 A

18 V dc 3.6 A

GNT60-24

24 V dc 2.1 A

24 V dc 1.9 A

24 V dc 2.7 A

GNT60-48

48 V dc 1.35 A

48 V dc 1.25 A

48 V dc 1.35 A