beautypg.com

Seaward 50 PrimeTest User Manual

Page 15

background image

PrimeTest 50

Operating Instructions

13

x

If the Insulation Resistance is lower than the
acceptable limit a cross is placed next to the
Insulation enunciator and the test sequence is
halted.

x

If the Insulation Resistance is greater than the
acceptable limit a tick is placed next to the
Insulation enunciator.

x

The unit will proceed with the wiring test,
checking the live and neutral conductors for
short or open circuits or reversed connections.

x

If the wiring is correct a tick is placed next to the
cord enunciator and a pass is indicated for the
sequence.