Operating Instructions

9

In the event of a failure, disconnect the PrimeTest 50
from the supply and rectify the fault.

Do not leave the PrimeTest 50 permanently
connected to a mains supply.

The auto switch-off function is disabled
when the PrimeTest 50 is connected to a
live mains socket. The unit will beep
continually after 3 minutes to remind the
user to disconnect from the mains socket.

3.2

Testing a Class I Appliance

x

Visually inspect the appliance and mains cord for
signs of damage.

x

If the appliance passes a visual inspection
proceed with the electrical tests.

x

Plug the earth test lead into the 4mm socket
on the PrimeTest 50 end panel.

x

Plug the appliance into the PrimeTest 50 front
panel mains socket

x

Connect the earth test probe to an exposed
metal part on the appliance.

x

If the Appliance under test has an ON/OFF
switch, make sure it is in the ON position.

x

Press the Class I test key

x

The PrimeTest 50 will now test the continuity of
the protective earth.

x

If the measured value is outside acceptable
limits a cross is placed next to the Earth Cont
enunciator, a FAIL is indicated and the test
sequence is halted.