Quantum Composers 9500 Plus Series User Manual
Page 37

33
- Within the Gate/Trig menu
EXTin:
Select Gated mode.
Level:
Set the gate threshold level.
Gate:
Select active high or active low.
Pressing the RUN/STOP key will arm the unit. Once the unit is armed, it will
begin generating pulses whenever the external trigger input is in the active state.
When the gate is in the active state, the system timer is reset. Pulses that have
already started when the gate enters the inactive state will continue until the
pulse is complete. Pressing the RUN/STOP key a second time will disarm the
unit.