Technical specifications – OWON ODP Series User Manual
Page 31

8.Technical Specifications
26
8. Technical Specifications
The specifications below are based on the instrument having run for at least 30 minutes
continuously under the specified operating temperature.
Channel 1/Channel 2
Fixed 5V
DC
Output Ratings
Voltage
Independent/Parallel
0~30V
5V
Series
0~60V
Plus-minus
-30V~30V
Current
ODP3032
Independent/Series
/Plus-minus
0~3A
3A
Parallel
0~6A
ODP3052
Independent/Series
/Plus-minus
0~5A
5A
Parallel
0~10A
Line Regulation
CV
≤0.01% + 3mV
≤3mV
CC
≤0.1% + 3mA
Load Regulation
CV
≤0.01% + 3mV
≤0.1%+3mV
CC
≤0.2% + 3mA
Noise and Ripple
(20Hz~7MHz)
CV
≤300
μVrms / 2 mVpp
≤300
μVrms
/ 2 mVpp
CC
≤3mArms
Settings Resolution
Voltage
1mV
None
Current
1mA
None
Settings Accuracy
(25℃±5℃)
Voltage
≤0.05% + 3mV
None
Current
≤0.1% + 3mA
None
Read Back
Resolution
Voltage
1mV (<10V)
10mV (≥10V)
None
Current
1mA
None
Read Back Accuracy
(25℃±5℃)
Voltage
≤0.05% + 3 digits
None
Current
≤0.1% + 3 digits
None