2.4.2 IEEE Bus Terminator Selection

SW3-6 through SW3-8 set the IEEE bus terminators used for data sent

or received by the Serial488A. EOI, a line used to signal the end of a
multiple character bus transfer, may also be enabled. If enabled, EOI is
asserted when the last selected bus terminator is sent. Factory default is LF
Only with EOI disabled.

SW3 View for IEEE Bus Terminator Selection

OPEN

1 2 3 4 5 6 7 8

DOT

Switch

Side
View

OPEN

1 2 3 4 5 6 7 8

CR Only

LF-CR

OPEN

1 2 3 4 5 6 7 8

OPEN

1 2 3 4 5 6 7 8

LF Only

CR-LF

OPEN

1 2 3 4 5 6 7 8

OPEN

1 2 3 4 5 6 7 8

EOI Disabled

EOI Enabled

2.5 Mode Selection

SW2-1 sets the major operating mode of the Serial488A. The IEEE Controller

(RS-232 to IEEE Converter) mode allows a serial host device to send data to a single
IEEE bus peripheral. Applications include interfacing a listen-only or addressable
IEEE printer/plotter to a serial printer port. Refer to Section 3 for more detailed
information on the controller mode of operation.