Two-wire configuration, Two-wire, single-sensor, Two-wire, two sensor – Measurement Computing WLS-TEMP User Manual
Page 15

WLS-TEMP Specifications
15
Two-wire configuration
The easiest way to connect an RTD sensor or thermistor to the WLS-TEMP is with a two-wire configuration,
since it requires the fewest connections to the sensor. With this method, the two wires that provide the RTD
sensor with its excitation current also measure the voltage across the sensor.
Since RTDs exhibit a low nominal resistance, the lead wire resistance can affect measurement accuracy. For
example, connecting lead wires that have a resistance of 1 Ω (0.5 Ω each lead) to a 100 Ω platinum RTD results
in a 1% measurement error.
With a two-wire configuration, you can connect either one sensor per channel pair, or two sensors per channel
pair.
Two-wire, single-sensor
A two-wire, single-sensor measurement configuration is shown in Figure 4.
Figure 4. Two-wire, single RTD or thermistor sensor measurement configuration
When you select a two-wire, single-sensor configuration with InstaCal, connections to C#H and C#L are made
internally.
Two-wire, two sensor
A two-wire, two-sensor measurement configuration is shown in Figure 5.
Figure 5. Two-wire, two RTD or thermistor sensors measurement configuration
When you select a two-wire, two-sensor configuration with InstaCal, connections to C#H (first sensor) and
C#H/C#L (second sensor) are made internally.
When configured for two-wire mode, connect both sensors to obtain proper measurements.