Accuracy, Stability – Measurement Computing USB-2404-UI User Manual
Page 17

USB-2404-UI User's Guide
Specifications
17
Current
±25 mA
±25 mA
4-wire and 2-wire resistance
10 k
Ω, 1 kΩ
10.5 k
Ω, 1.05 kΩ
Thermocouple
±125 mV
±125 mV
4-wire and 3-wire RTD
Pt 1000, Pt 100
5.05 k
Ω, 505 Ω
Quarter-bridge
350
Ω, 120 Ω
390
Ω, 150 Ω
Half-bridge
±500 mV/V
±500 mV/V
Full-bridge
±62.5 mV/V, ±7.8 mV/V
±62.5 mV/V, ±7.8125 mV/V
Accuracy
Table 3. Accuracy specifications
Mode, range
Gain error (percent of reading)
Offset error (ppm of range)
Typical 25 ºC, ±5 ºC, –40 to 70 ºC maximum
Voltage, ±60 V
±0.3, ±0.4
±20, ±50
Voltage, ±15 V
±0.3, ±0.4
±60, ±180
Voltage, ±4 V
±0.3, ±0.4
±240, ±720
Voltage, ±1 V
±0.1, ±0.18
±15, ±45
Voltage/Thermocouple, ±125 mV
±0.1, ±0.18
±120, ±360
Current, ±25 mA
±0.1, ±0.6
±30, ±100
4-wire and 2-wire (note 1) resistance,
10 k
Ω
±0.1, ±0.5
±120, ±320
4-wire and 2-wire (note 1) resistance,
1 k
Ω
±0.1, ±0.5
±1200, ±3200
4-wire and 3-wire RTD, Pt 1000
±0.1, ±0.5
±240, ±640
4-wire and 3-wire RTD, Pt 100
±0.1, ±0.5
±2400, ±6400
Quarter-bridge, 350
Ω
±0.1, ±0.5
±2400, ±6400
Quarter-bridge, 120
Ω
±0.1, ±0.5
±2400, ±6400
Half-bridge, ±500 mV/V
±0.03, ±0.07
±300, ±450
Full-bridge, ±62.5 mV/V
±0.03, ±0.08
±300, ±1000
Full-bridge, ±7.8 mV/V
±0.03, ±0.08
±2200, ±8000
Cold-junction compensation sensor
accuracy
±1 ºC, typical
Note 1:
2-wire resistance mode accuracy depends on the lead wire resistance. This table assumes 0
Ω of lead
wire resistance.
Stability
Table 4. Stability specifications
Measurement conditions
Gain drift (ppm of reading/ ºC)
Offset drift (ppm of range/ ºC)
Voltage, ±60 V
±20
±0.2
Voltage, ±15 V
±20
±0.8
Voltage, ±4 V
±20
±3.2
Voltage, ±1 V
±10
±0.2
Voltage/thermocouple, ±125 mV
±10
±1.6
Current, ±25 mA
±15
±0.4
4-wire and 2-wire resistance, 10 k
Ω
±15
±3
4-wire and 2-wire resistance, 1 k
Ω
±15
±30
4-wire and 3-wire RTD, Pt 1000
±15
±6
4-wire and 3-wire RTD, Pt 100
±15
±60
Quarter-bridge, 350
Ω
±15
±120
Quarter-bridge, 120
Ω
±15
±240
Half-bridge, ±500 mV/V
±3
±20