beautypg.com

Specifications, Analog input, Input mode ranges – Measurement Computing USB-2404-UI User Manual

Page 16

background image

16

Chapter 4

Specifications

All specifications are subject to change without notice.
Typical for the range 0 to 60 °C unless otherwise noted.

Analog input

Table 1. Analog input specifications

Parameter

Conditions

Specification

Number of channels

4

A/D converter resolution

24-bit

A/D converter type

Delta-Sigma with analog pre-filtering

Sampling mode

Simultaneous

TEDS sensor type supported

IEEE 1451.4 TEDS Class II (interface)

Input mode

 Voltage
 Current
 Resistance (4-wire and 2-wire)
 RTD (3-wire and 4-wire)
 Thermocouple
 Quarter-bridge
 Half-bridge
 Full-bridge

Conversion time
(No channels in TC mode)

High speed

10 ms for all channels

Best 60 Hz rejection

110 ms for all channels

Best 50 Hz rejection

130 ms for all channels

High resolution

500 ms for all channels

Conversion time
(One or more channels in TC mode)

High speed

20 ms for all channels

Best 60 Hz rejection

120 ms for all channels

Best 50 Hz rejection

140 ms for all channels

High resolution

510 ms for all channels

Overvoltage protection

Terminals 1 and 2

±30 V

Terminals 3 through 6, across any
combination

±60 V

Input impedance

Voltage mode
(±60 V, ±15 V, ±4 V)

1 M

Current mode

<40

All other modes

>1 GΩ

Input bias current

<1 nA

Integral non-linearity (INL)

±15 ppm

Common mode rejection ratio
(CMRR)

>100 dB

Normal mode rejection ratio
(NMRR)

Best 60 Hz rejection

90 dB at 60 Hz

Best 50 Hz rejection

80 dB at 50 Hz

High resolution

65 dB at 50 Hz and 60 Hz

Input mode ranges

Table 2. Input mode range specifications

Input mode

Nominal range(s)

Actual range(s)

Voltage

±60 V, ±15 V, ±4 V, ±1 V, ±125 mV

±60 V, ±15 V, ±4 V, ±1 V, ±125 mV