USB-1208FS-Plus User's Guide

Functional Details

12

Signal connections

Analog input

You can connect up to eight analog input connections to the screw terminal bank containing pins 1 to 20.

You can configure the analog input channels as eight single-ended channels (

CH0

IN

through

CH7

IN)

or four

differential channels (

CH0

IN

HI/LO

through

CH3

IN HI/LO

). When configured for differential mode, each

analog input has 12-bit resolution. When configured for single-ended mode, each analog input has 11-bit
resolution, due to restrictions imposed by the A/D converter.

Single-ended configuration

When configured for single-ended input mode, the input signal is referenced to signal ground (GND) and
delivered through two wires:

 Connect the wire carrying the signal to be measured to

CH# IN

.

 Connect the second wire to

AGND

.

The input range for single-ended mode is ±10 V. The single-ended mode pinout is shown in Figure 3 on page
11.

Differential configuration

When configured for differential input mode, the input signal is measured with respect to the low input and
delivered through three wires:

 Connect the wire carrying the signal to be measured to

CH# IN HI

.

 Connect the wire carrying the reference signal to

CH# IN LO

.

 Connect the third wire to

GND

.

The differential mode pinout is shown in Figure 4 on page 11.

Note

: To perform a single-ended measurement using differential channels, connect the signal to

CH# IN HI

and

ground the associated

CH# IN LO

input.

A low-noise precision programmable gain amplifier (PGA) is available on differential channels to provide gains
of up to 20 and a dynamic range of up to 12-bits. Differential mode input voltage ranges are ±20 V, ±10 V,
±5 V, ±4 V, ±2.5 V, ±2.0 V, ±1.25 V, and ±1.0 V.

In differential mode, the following two requirements must be met for linear operation:

 Any analog input must remain in the −10V to +20V range with respect to ground at all times.
 The maximum differential voltage on any given analog input pair must remain within the selected voltage

range.

The input [common-mode voltage + signal] of the differential channel must be in the −10 V to +20 V range in
order to yield a useful result.