Note 1:

Input current is a function of applied voltage on the analog input channels. For a given input
voltage, V

in

, the input leakage is approximately equal to (8.181 * V

in

– 12) µA.

Note 2:

Maximum throughput when scanning is machine dependent.

Note 3:

The AD7870 converter only returns 11 bits (0 to 2,047 codes) in single-ended mode.

Accuracy

Table 2. Accuracy, differential mode

Range

Accuracy (LSB)

±20 V

5.1

±10 V

6.1

±5 V

8.1

±4 V

9.1

±2.5 V

12.1

±2 V

14.1

±1.25 V

20.1

±1 V

24.1

Table 3. Accuracy, single-ended mode

Range

Accuracy (LSB)

±10 V

4.0

Table 4. Accuracy components, differential mode – all values are (±)

Range

% of Reading

Gain Error at FS (mV)

Offset (mV)

Accuracy at FS (mV)

±20 V

0.2

40

9.766

49.766

±10 V

0.2

20

9.766

29.766

±5 V

0.2

10

9.766

19.766

±4 V

0.2

8

9.766

17.766

±2.5 V

0.2

5

9.766

14.766

±2 V

0.2

4

9.766

13.766

±1.25 V

0.2

2.5

9.766

12.266

±1 V

0.2

2

9.766

11.766

Table 5. Accuracy components, single-ended mode – all values are (±)

Range

% of Reading

Gain Error at FS (mV)

Offset (mV)

Accuracy at FS (mV)

±10 V

0.2

20

19.531