beautypg.com

The autolog function – Delta Electronics SS1-UM-1.05 User Manual

Page 47

background image

SunScan User Manual v 1.05

Measurement options

••••

47

The Autolog function

You can set up the SunScan system to operate automatically in any of its modes, just
as if you were pushing the read, store and average buttons at regular intervals.

Note: don't confuse Autolog with logging the probe as a Linear Quantum sensor, or
with logging the Beam Fraction sensor as an independent sensor. These applications
are dealt with in the Appendices.

From

Menu, Utils, Autolog, Enter

you will arrive in the

Automatic Logging

setup

screen:

The

Start

time can be set to a time up to 24 hours before the current time! You might

do this for example if you wanted to set up averaging on the hour with sampling at
one minute intervals but the system time had just gone past the hour.

When you are satisfied with your choices, press

Enter

to accept the screen and

initiate the logging process. After a few minutes the Workabout will sleep and the
screen will go blank. You can safely press

Off

to blank the screen immediately (it

will not stop the Autolog process). From time to time when a reading is logged, the
screen will briefly show a

Logging in progress

message.

At any time, if you want to check the status of the Autolog, press

On/Esc

to access

the screen on the left.

The timing of the next reading (and the Plot and Sample number) will be indicated,
with the current system time. You can either select

Sleep, Enter

(to continue) or

Cancel, Enter

to cancel (stop) the logging session from this screen. If you select

Cancel

you must confirm that you want to stop logging by pressing

On/Esc

.

System time in the heading is
constantly updated

You will be prompted with a current
Start time, and 6 hours later Stop time
and date. Edit these appropriately.

Change the read interval to
whatever sampling interval you
want (hh:mm:ss).

If you leave the Average interval at 0
you will not get average summaries.
Any other setting (hh:mm:ss) will
insert average summaries into the data
file at the specified intervals.

Nighttime is defined as when the
sun is more than 6

°°°°

below the

horizon. SunScan computes this
for you automatically.

This is the Autolog setup screen.
Use the

↑↑↑↑↓↓↓↓

keys to select the

choices; edit the prompted values
or use the

keys to toggle

through options presented.