LTL2000S(Q) Retroreflectometer

34

Command set

Command


Parameter


Response


Meaning

Rl Status Code 0 00000000
2002 Aug 12 13:53:19
LTL2000SQ v1.0 s/n:n

Rl status code
Time for Rl measurement
Instrument Id

OTn None

59
180

Off Timer = 300 sec.
Off Func: Off
Off Timer: = 180 sec.

Automatic turn off time when not in use. Range 60 to
600 sec. Disable for time < 60 sec

PV

None

V_PMT = 644 Volt

Voltage on PMT

QCn

None
123

QdF: 0.1366
Cal Qd? [Y/N]
Cal Qd!
Qd Factor = 0.7564
Qd Normal = 123

Query Qd calc factor
Calibrate Qd to 123 Y/N response to start
Calibrating Qd
Calculated calc factor
New active Qd normal

QEn None

F
T

Qd On
Qd Off
Qd On

Query Qd function
Set Qd function Off
Set Qd function On

RCn None

123

RlF: 2.038
Cal Rl [Y/N]
Cal Rl
Rl Factor = 2.032
Rl Normal = 123

Query Rl Calc factor
Calibrate Rl to 123 Y/N response to start
Calibrating Rl
Calculated calc factor
New active Rl normal

REn None

F
T

Rl On
Rl Off
Rl On

Querry Rl function
Set Rl function Off
Set Rl function On

RZ None Zero

Adjust?

[Y/N]

Zero Rl
*** Zero Reading ***
Lamp On: 12
Lamp Off: 10
New Rl_Sy_Leak: 2

Perform Rl zero adjust with Y/N confirmation
Measuring zero signal
Result from zero reading
Signal with lamp on
Signal with lamp off
New SyLeak signal

SD None

Rl Status Code 0 00000000

Qd Status Code: 0 00000000

Status code and bit pattern
if status not zero then the status code is decoded

SNn None



TEST2

Seq ID: DELTA
Meas in Seq: 3

Seq ID: TEST2
Meas in Seq: 0

Querry Sequence number
Measurements with this ID

Change ID to TEST2
No measurements with this ID

TIn

None
10 02 25

2002 Aug 12 13:59:37
2002 Aug 12 10:02:25

Real time clock Date and Time
New Time (Date unchanged)