LTL2000S(Q) Retroreflectometer

11

SECTION 2

GENERAL INFORMATION


Rl Measurement

The Rl part of the LTL2000S(Q) Retroreflectometer measures the Rl (coefficient of retro re-
flected luminance) parameter. The Rl parameter represents the brightness of the road mark-
ings seen by drivers of motor vehicles by headlight illumination.

The illumination angle is 1.24

° and the observation angle is 2.29° simulating a drivers view-

ing distance of 30 metres at an eye height of 1.2 metres. The observation area is app. 45 mm.
x 200 mm.

The optical specification complies with EN1436. The spectral sensitivity is eye matched us-
ing SMART optical filter designed and produced by DELTA. The functional layout of the
LTL2000S(Q) Rl optical system is shown in figure 3 below.

A very efficient electronic stray light compensation eliminates false signals from daylight.

101

102

103

104

107

108 109

105

106

112

111

110


Figure 3: LTL2000S(Q) RL optical system layout

101 Light source
102 Receiver and spectral matching

filter

103 Illumination aperture stop
104 Observation aperture stop
105 Illumination beam
106 Observation beam
107 Collimating lens
108 Mirror
109 Beam splitter
110 Road marking surface
111 Illumination field
112 Observation field