LTL2000S(Q) Retroreflectometer

7

The display shows:
Step 1

Step 2

Step 3

Step 4

Step 5




Step 6

Step 7

Step 8





To verify the calibration do a Rl measurement on the normal before removing the normal in
step 8.
To document the calibration press the key to print.

In the example above the Rl function is now calibrated to a Rl value of 150.
If an error is detected during the calibration procedure then the display will indicate this.

Note:

Mount Light Trap

when Ready

Zeroing
Please Wait

Zero OK
Zero: 0.05%

Mount Rl Normal

when ready

Calibrating

Please Wait

Rl Normal 150

= when OK

Calibration Done
Remove Normal

Rl Normal 150

= OK

to Edit

Wait

• Avoid calibration in direct sunlight

• Store the reference calibration unit in a dry and clean environment.