beautypg.com

Ivoclar Vivadent EP3000-G2 User Manual

Page 33

background image

33

Please contact the Ivoclar Vivadent After Sales Service, if one of the
following error messages is being displayed:

25, 29

43, 44, 45, 46, 47, 48

54, 56

103, 107

143, 144, 145, 146, 147, 148

700, 701, 703, 704, 705, 706, 707

1010, 1011, 1012, 1013, 1014, 1015, 1016

1024, 1025, 1026, 1028

1143, 1144, 1145, 1146, 1147, 1148

1202, 1203, 1204, 1205, 1206, 1207

1400, 1401, 1402

1500

*

Furnace head opens when this error occurs.

** A program in progress is stopped.
*** The error cannot be acknowledged; the programs cannot be started.

Error /

Hint

No.

Continua-

tion pos-

sible

Error

Error Message Text

520

Error muffle crack

CDS Crack Detection System has been activated. The program has been aborted and the press

plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.

Please check your restorations before you continue your working progress.

521

Error muffle crack

CDS Crack Detection System has been activated. The program has been aborted and the press

plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.

Please check your restorations before you continue your working progress.

522

Error muffle crack

CDS Crack Detection System has been activated. The program has been aborted and the press

plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.

Please check your restorations before you continue your working progress.

530

Error during logging of press

program data

An error has occurred during logging of press program data. The storage medium might be full.

700

Supply voltage outside the

acceptable range

The supply voltage is outside the acceptable range. Check the supply voltage.

701

***

no

Start-up aborted due to an error The self-test of the furnace was interrupted by an error. It is not possible to work with the furnace!

Switch the furnace off and on again, once the error has been rectified.

702

Brief power failure during a

program in progress

A firing program in progress was interrupted by a brief power failure. The program is continued!

704

Power failure during an over-

night program in progress

An overnight program in process was interrupted by a power failure. The program is continued!

707

Incorrect supply voltage

The furnace is operated with the incorrect supply voltage. Make sure that the furnace is operated

with the supply voltage indicated on the rating plate.

800

Final vacuum value not reached

The required final vacuum value cannot be reached. Check the vacuum pump.

801

Vacuum drop

An unacceptable vacuum drop has occurred

802

The vacuum does not increase

(self-test)

No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no

contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump con-

nected? Is the fuse F1 o.k.?

1310

Calibration reminder

Some time has passed since the last calibration procedure. Calibrate the furance soon.

1311

Calibration reminder – press

cycles

Some press cycles has been done since the last calibration procedure. Calibrate the furance soon.

1312

Dehumidification reminder

Some time has passed since the last dehumidification. Conduct a dehumidification in the near future.

1550

Change operation mode

The operation mode has been changed! Make sure that the furnace runs at the new stand-by tem-

perature long enough before you start a program.

1700

Service note

Two years have passed or the heating muffle has been in use for more than 1200 firing hours since

the last inspection of the furnace. Ivoclar Vivadent therefore recommends that your furnace should

be inspected. Please refer to the Equipment Service Passport or the Operating Instructions for

further information. The interval until the next appearance of the note can be set in the Settings.