Ivoclar Vivadent EP3000-G2 User Manual
Page 33
33
Please contact the Ivoclar Vivadent After Sales Service, if one of the
following error messages is being displayed:
25, 29
43, 44, 45, 46, 47, 48
54, 56
103, 107
143, 144, 145, 146, 147, 148
700, 701, 703, 704, 705, 706, 707
1010, 1011, 1012, 1013, 1014, 1015, 1016
1024, 1025, 1026, 1028
1143, 1144, 1145, 1146, 1147, 1148
1202, 1203, 1204, 1205, 1206, 1207
1400, 1401, 1402
1500
*
Furnace head opens when this error occurs.
** A program in progress is stopped.
*** The error cannot be acknowledged; the programs cannot be started.
Error /
Hint
No.
Continua-
tion pos-
sible
Error
Error Message Text
520
Error muffle crack
CDS Crack Detection System has been activated. The program has been aborted and the press
plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.
Please check your restorations before you continue your working progress.
521
Error muffle crack
CDS Crack Detection System has been activated. The program has been aborted and the press
plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.
Please check your restorations before you continue your working progress.
522
Error muffle crack
CDS Crack Detection System has been activated. The program has been aborted and the press
plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.
Please check your restorations before you continue your working progress.
530
Error during logging of press
program data
An error has occurred during logging of press program data. The storage medium might be full.
700
Supply voltage outside the
acceptable range
The supply voltage is outside the acceptable range. Check the supply voltage.
701
***
no
Start-up aborted due to an error The self-test of the furnace was interrupted by an error. It is not possible to work with the furnace!
Switch the furnace off and on again, once the error has been rectified.
702
Brief power failure during a
program in progress
A firing program in progress was interrupted by a brief power failure. The program is continued!
704
Power failure during an over-
night program in progress
An overnight program in process was interrupted by a power failure. The program is continued!
707
Incorrect supply voltage
The furnace is operated with the incorrect supply voltage. Make sure that the furnace is operated
with the supply voltage indicated on the rating plate.
800
Final vacuum value not reached
The required final vacuum value cannot be reached. Check the vacuum pump.
801
Vacuum drop
An unacceptable vacuum drop has occurred
802
The vacuum does not increase
(self-test)
No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump con-
nected? Is the fuse F1 o.k.?
1310
Calibration reminder
Some time has passed since the last calibration procedure. Calibrate the furance soon.
1311
Calibration reminder – press
cycles
Some press cycles has been done since the last calibration procedure. Calibrate the furance soon.
1312
Dehumidification reminder
Some time has passed since the last dehumidification. Conduct a dehumidification in the near future.
1550
Change operation mode
The operation mode has been changed! Make sure that the furnace runs at the new stand-by tem-
perature long enough before you start a program.
1700
Service note
Two years have passed or the heating muffle has been in use for more than 1200 firing hours since
the last inspection of the furnace. Ivoclar Vivadent therefore recommends that your furnace should
be inspected. Please refer to the Equipment Service Passport or the Operating Instructions for
further information. The interval until the next appearance of the note can be set in the Settings.