Guralp Systems CMG-EDU User Manual
Page 33

Operator's guide
You will need to RE-BOOT after issuing this command for it to take
effect.
5.2.2
.IDS
Syntax: .IDS
Displays the System ID and stream IDs which are used to stamp all
outgoing data packets, in the form
system-id stream-id sensor-type ok
where system-id is the name of the instrument (as reported in
Scream!'s instrument list) and stream-id is the string which begins
every data stream.
This command is in the extended dictionary; to use it, first issue the
command ok-1 and finish with [seal].
5.2.3
BAUD
Syntax: 0 baud-rate BAUD
Sets the baud rate for the serial port on the CD24E1, in bytes per
second. The CD24E1 has a single port, numbered 0. For example,
0 19200 BAUD
This will reset a standard CD24E1 to its default configuration.
The allowable values for baud-rate are 4800, 7200, 9600, 14400,
19200, 38400, 57600 and 1152 (n.b.), representing 115200.
You will need to RE-BOOT after issuing this command for it to take
effect.
5.2.4
GO
Syntax: GO
Exits the terminal session and begins sending data. In Scream!, closing
the terminal window has the same effect as this command.
If you have made configuration changes, you may need to RE-BOOT
the digitizer instead of using this command.
October 2005
33