B-9
Table B-2. Factory Sequences (MPS450 and Marq III) (cont.)
Dual
Keys Sequence
Segments/Selections
Run
Times
Serial
Command
Set ECG type to adult
0:00:00
ADULT
ECG/NSR BPM 80
0:30:00
NSR60
PVCs 6/Min
0:00:00
PVC6
ECG Muscular Artifact ON
0:01:00
EAMSC
PVCs 12/Min
0:01:00
PVC12
ECG Muscular Artifact OFF
0:29:00
EAOFF
ECG/NSR BPM 80
1:30:00
NSR80
PVCs 12/Min
0:01:00
PVC12
ECG/NSR BPM 80
1:29:00
NSR80
ECG/NSR BPM 60
4:00:00
NSR60
Pair PVCs (Couplet)
0:01:00
PAIR
Run 5 PVCs
0:01:00
RUN5
ECG/NSR BPM 80
0:58:00
NSR80
Q
L
P
Holter ECG
Series
REPEAT
Set ECG type to adult
0:00:00
ADULT
Demand 1
0:10:00
DOS
Pacer Non-capture
0:09:00
NCA
Pacer Non-function 0:01:00
NFU
Demand 2
0:10:00
DFS
Q
M
B
Pacemaker
Series
REPEAT
Set ECG type to pediatric
0:00:00
PEDS
ECG/NSR 140 BPM
0:00:10
NSR140
Set respiration base to 500
• 0:00:00
RB500
Set respiration delta to 1.0
• 0:00:00
RO1.0
Set respiration rate to 60BrPM
0:00:00
RR60
ECG/NSR BPM 60
0:00:00
NSR60
Set Apnea to 22 seconds
0:00:00
A22
ECG/NSR BPM 60
0:00:06
NSR60
ECG/NSR BPM 80
0:00:06
NSR80
ECG/NSR BPM 120
0:00:06
NSR120
Q
M
C
Peds/Apnea
/Brady
ECG/NSR BPM 140
0:00:06
NSR140