Fluke Biomedical 945A User Manual
Page 81

Preamplifier Interface Board 977-210-10 Switch Settings
Switch
Normal Op. Switch Function
SW3-1
ON
0/OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)
SW3-2
OFF
0/OFF Selects alarm inhibit, SW3-3 must be on for this state to be true.
1/ON Selects alarm enable
SW3-3
OFF
0/OFF Selects Rate Mode (normal) • 1/ON Selects Maintenance Mode
SW3-4
OFF
0/OFF forces 977 analog output to 0 V (low scale)
(SW3-3 must be on for state to be true)
1/ON forces 977 analog output to 10 V
(SW3-3 must be on for state to be true)
NOTE
The above switches are located in a 4 position dip switch on the
977 interface board (977-210-10).
Jumper Positions for Preamplifier Interface Board 977-210-10
Jumper Normal
Setting
Function Alternate
Setting
Function
JMP1
A to B
Enables Victoreen Loop (normal)
B to C
Enables RS232C
JMP2
N/A
Deleted, Not Used
N/A
Deleted, Not Used
JMP3
A to B
Enables Alarm Output -
Non-Failsafe
B to C
Enables Alarm Output
Failsafe
JMP4
A to B
Enables 0 - 10 Vdc Analog Output
(normal)
B to C
Disables Analog Output (stays at
10 Vdc)
JMP5
B to C
Enables HV Output (normal)
A to B
Disables HV Output
Jumper Positions for Preamplifier Electrometer Board 977-200-15
Jumper Normal
Setting
Function Cal.
Position
Function
JMP1
A to B
Normal Operation
B to C
PGA Offset Adjust *
* Programmable Gain Amplifier
Model 946A-200 Option Jumpers
Jumper Function
Norm
Position
Operation
JP3-1,
Not Defined
in
Not Applicable
JP3-2
Not Defined
in
Not Applicable
JP3-3
Alarm Reset Select
in
Manual reset with jumper installed
Auto reset with jumper removed
JP3-4
Fail Alarm Enable
in
Fail alarm enabled with jumper in
Fail alarm disabled with jumper out
JP3-5
Check Source Alarm Mute in
Alarms disabled during C/S with
jumper in
Alarms enabled during C/S with
jumper out
8-5