Fluke Biomedical 945A User Manual
Page 56

Relay Control Register (Write Only)
The Relay Control Register (U44) is an 8 bit register with clear, and responds to address 400C. Upon
initial power-up, the system reset signal sets all outputs low. Data written into U44 remains at the outputs
until a reset occurs or new data is written. Table 4-15 lists relay control write functions.
Table 4-15. Relay Control Register Write Functions
Data
Bit
Function
D0
Fail Relay and Indicator
D1
Check Source Relay
D2
Warn
Relay
D3
Alarm
Relay
D4
Rate-of-rise Relay (not used)
D5
Remote
Indicator
With the exception of the fail bit (D0), all outputs are applied to U48, which is an inverting open collector
driver. The outputs of U48, including fail, drive (via the J2 connector) mechanical relays located on the
relay board. The fail bit is used as an input to U47-3 which, when set high-low-high once per second,
causes output U47-6 to remain low. Should this high-low-high sequence fail to occur (under MPU
control), U47 will time out and set U47-6 high, causing the fail indicator on the panel and the fail relay to
de-activate.
Bargraph (Write Only)
The bargraph addresses are decoded by U5, which is a 1 of 8 decoder. Control line inputs to U5 are
R/W, Write 02, and Bargraph. When the bargraph address is selected (4020-4025), U5 further decodes
the addresses listed in Table 4-16.
Table 4-16. Bargraph Write Address
Address State
Hex Address
Function
Driver
A3 A2 A1
0 0 0
4020
Green
1-10
U55,
U56
0 0 1
4021
Green
11-20
U57,
U58
0 1 0
4022
Green
21-24
U59
0 1 1
4023
Red
1-10
U50,
U51
1 0 0
4024
Red
11-20
U52,
U53
1 0 1
4025
Red
21-24
U54
4-32