Section 6 - Calibration
Calibration
The 977 Ion Chamber Detector is factory calibrated on a 137Cs/60Co range. A ± 15%, 10 point NIST
traceable factory calibration is performed to obtain a calibration factor for the detector. The calibration
factor is stated on the factory calibration data sheet, and is adsjusted via set-point 5 on the 946A UDR. A
field test source, P/N 848-5, is available as an option to verify detector operation. The
detector/preamplifier pair should be returned to Fluke Biomedical RMS for recalibration whenever
performance degredation is observed.
Electronic Adjustments
Table 6-1 lists the electronic adjustments for the 946A UDR and 977-210 preamplifier. The procedures
for making these adjustments are contained in the factory Calibration Procedure listed below. Note that
test firmware is required to perform the electrical adjustments. Refer to Section 7 for a description of user
permitted adjustments.
Table 6-1. Electronic Adjustments
Adjustment
(UDR)
Purpose
Tolerance
VR1
Zero Adjust / 0 - 10 Volt Analog Output
0.000 Vdc ±5 mV
VR2
Gain Adjust / 0 - 10 Volt Analog Output
±.5% of full scale
VR4
20 mA Adjust / 4 - 20 mA Volt Analog Output #1
±.5% of full scale
VR5
20 mA Adjust / 4 - 20 mA Volt Analog Output #2
±.5% of full scale
VR6
4 mA Adjust / 4 - 20 mA Volt Analog Output #2
±.5% of full scale
VR7
4 mA Adjust / 4 - 20 mA Volt Analog Output #1
±.5% of full scale
VR13
Write Cycle Clock Adjust
±5 ns
Adjustment
(977)
Purpose
R11
Programmable Gain Amplifier Offset Adjustment
± 2 mV
R12
Auto Zero Set-Up Adjustment
± 10 mV
HV Supply R118
High Voltage Adjust
± 20 V
R53
Analog Output Adjustment
± 10 mV
Note
Refer to the drawings in Appendix B to locate adjustment and/or test points, if required.
WARNING
Sections of this circuit board contain voltages in excess of 500
volts. Use extreme care when making any adjustments.
Use extreme care when making high voltage measurements directly
at J6 or TP3.
The follow factory calibration and test procedures are included for reference.
Document
Description
Cal-ION2
Factory Calibration Procedure for the 977 Detector
TP977-201/210
Test Procedure for 977 Series
LT946/977
Factory Loop Test for 945 System
TP946A-100-3
Factory Test Procedure for 946A-200
MB95-1004
Field Cable Termination for Replacement 977-210 Detector
TP-977 Cable
Factory Test Procedure for 977-210-66 cables
6-1