Section 3 - installation -1, Nstallation, Detector mounting -1 udr mounting -1 – Fluke Biomedical 945A User Manual
Page 19

Section 3 - Installation
Installation
Installation consists of mounting the equipment, making the required electrical connections, and entering 
the desired set points. Installation drawings are provided in Appendix B of this manual. 
CAUTION
Remove all power prior to installing the UDR.
Detector Mounting
Refer to Engineering Drawing GEL 977-210, provided in Appendix B, for detector and preamplifier 
mounting dimensions, weights and recommended fasteners. 
UDR Mounting
The UDR is a self contained unit which can be mounted in four different ways.
CAUTION
When installing the UDR in enclosed panels or cabinets, ensure that the maximum
operating temperature (122° F) is not exceeded. The total heat load should be 
calculated to determine whether cooling by natural convection or forced 
ventilation (i.e. fans) is required. The heat load for a single UDR is approximately 
96 BTU/hr. 
 
When more than four units are to be mounted in a cabinet, it is recommended that 
the center position in each chassis be left blank to facilitate the flow of cooling air 
through the cabinet. 
NOTE
For seismic applications, the rear of the chassis must be supported, or the sides of 
the chassis must be fastened together. For further information, contact Fluke 
Biomedical RMS. 
1. The first mounting utilizes a 19 inch rack chassis adapter, Victoreen Model 948-1. This configuration
requires 3.5 inches of rack height and can be used to mount up to three UDRs side by side. A blank 
filler panel, Victoreen Model 948-2, is available to cover unused mounting positions. In multiple UDR 
installations, the center mounting bay should be fitted with a Model 948-2 blank panel to facilitate 
cooling air flow. 
2. The second mounting consists of a single channel table top enclosure, Victoreen Model 948-3.
Dimensions for the enclosure are 4.5 inches high by 6.8 inches wide by 14.9 inches deep. This 
enclosure is used primarily in a laboratory environment. Refer to drawing 948-3-5 for mounting 
dimensions. 
3. The third mounting consists of a panel mount enclosure, Victoreen Model 948-9. Cutout dimensions
for this enclosure are 3.75 inches high by 6.45 inches wide, the outside dimensions for the enclosure 
are 4.0 inches high by 6.7 inches wide by 13.7 inches deep. This mounting must also be rear 
supported in seismic applications. Refer to drawing 948-9-5 for mounting dimensions. 
3-1