Udr function switch / function -17, Rror, Odes – Fluke Biomedical 945A User Manual
Page 41

UDR Function Switch / Function
Table 4-7 shows the UDR Function Switch positions and their function.
Table 4-7. UDR Function Switch Positions/Function
Switch
Position
Function
0
Selects HIGH Alarm Setpoint
1
Selects WARN Alarm Setpoint
2
Selects Overrange Limit
3
Selects Underrange Limit
4
Selects Analog Full Scale
5
Selects Calibration Factor
7
Selects Analog Low Scale
8
Selects Auto Zero Display Mode
9
Selects Loop Test Mode
Error Codes
The error code for the 946A-200 UDR are listed in Table 4-8.
Table 4-8. Model 946A-200 Error Codes
Error Codes
Function
Reference Section
E0001
Negative display data
Setpoint Error Codes
E0002
Invalid setpoint value (s)
Setpoint Error Codes
E0007
Loop time out
Communication Loop Failure
E0008
Invalid analog scale values
Setpoint Error Codes
E0009
Loop test failure
Loop Test Mode
E0011*
High voltage failure
Fail Alarms
E0012*
Auto zero correction failure
Fail Alarms
* These error codes always indicate a detector failure.