Calibration – Fluke Biomedical ULT800 User Manual
Page 25

Ultrasound Transducer Leakage Current Tester
Maintenance
15
Calibration
Warning
Examine the calibration label on the back of the ULT800
prior to each use. Do not use a ULT800 with an expired
calibration label. A ULT800 without a calibration label or
with the anti-tamper case label broken is out of calibration.
A ULT800 that is out of calibration can cause excessive
leakage current exposure to the patient; risk of injury to
the patient could result.
Caution
Avoid dropping the ULT800 or allowing other mechanical
abuse that could cause a shift in the ULT800’s calibrated
settings.
The ULT800 requires yearly Fluke factory calibration, which uses
appropriate tools and reference instruments that are traceable to
the National Institute of Standards and Technology (NIST). Factory
calibration provides a calibration sticker on the back of the ULT800
to verify that the calibration was performed.
To locate a service center, visit the Fluke web site at
www.fluke.com, or contact Fluke at [email protected]. Call from
anywhere in the world at +1-425-446-5500 or call for service in the
USA at 1-888-99-FLUKE (1-888-993-5853.)