Specifications – Fluke Biomedical ULT800 User Manual
Page 16

ULT800
Users Guide
6
Specifications
Power ...................................................................... 9 V Alkaline Battery
No. of Measurements ............................................ Approx. 1000 measurements on a
single battery
Conductivity ........................................................... Pass limit is >250
μ
A
±
5 %
Leakage .................................................................. Pass limit is <185
μ
A
±
5 % and
>20
μ
A
±
5
μ
A
Dimensions ............................................................ 6.6 in x 3.7 in x 1.5 in
(17 cm x 19 cm 4 cm)
Weight ..................................................................... 12 oz (340 g)
Temperature
Operation ............................................................. 15
°
C to 40
°
C
Storage ................................................................ 15
°
C to 65
°
C
Relative Humidity ................................................... 90 % max.
Altitude ................................................................... 2000 m
IP Rating ................................................................. IEC 60529: IP20
Safety ...................................................................... IEC 61010-1: Pollution Degree 2
Electromagnetic Environment .............................. IEC 61326-1: Basic
Emissions Classification ...................................... IEC CISPR 11: Group 1, Class A.
Group 1 have intentionally generated
and/or use conductively coupled radio-
frequency energy which is necessary
for the internal functioning of the
equipment itself. Class A equipment is
suitable for use in non-domestic
locations and/or directly connected to
a low-voltage power supply network.
Electromagnetic Compatibility ............................. Applies to use in Korea only.Class A:
Equipment (Industrial Broadcasting &
Communication Equipment)
[1]
[1] This product meets requirements for industrial (Class A) electromagnetic wave
equipment and the seller or user should take notice of it. This equipment is intended