beautypg.com

Specifications – Fluke Biomedical ULT800 User Manual

Page 16

background image

ULT800

Users Guide

6

Specifications

Power ...................................................................... 9 V Alkaline Battery
No. of Measurements ............................................ Approx. 1000 measurements on a

single battery

Conductivity ........................................................... Pass limit is >250

μ

A

±

5 %

Leakage .................................................................. Pass limit is <185

μ

A

±

5 % and

>20

μ

A

±

5

μ

A

Dimensions ............................................................ 6.6 in x 3.7 in x 1.5 in

(17 cm x 19 cm 4 cm)

Weight ..................................................................... 12 oz (340 g)
Temperature

Operation ............................................................. 15

°

C to 40

°

C

Storage ................................................................ 15

°

C to 65

°

C

Relative Humidity ................................................... 90 % max.
Altitude ................................................................... 2000 m
IP Rating ................................................................. IEC 60529: IP20
Safety ...................................................................... IEC 61010-1: Pollution Degree 2
Electromagnetic Environment .............................. IEC 61326-1: Basic
Emissions Classification ...................................... IEC CISPR 11: Group 1, Class A.

Group 1 have intentionally generated
and/or use conductively coupled radio-
frequency energy which is necessary
for the internal functioning of the
equipment itself. Class A equipment is
suitable for use in non-domestic
locations and/or directly connected to
a low-voltage power supply network.

Electromagnetic Compatibility ............................. Applies to use in Korea only.Class A:

Equipment (Industrial Broadcasting &
Communication Equipment)

[1]

[1] This product meets requirements for industrial (Class A) electromagnetic wave

equipment and the seller or user should take notice of it. This equipment is intended
for use in business environments and is not to be used in homes.