Figure 2. disinfection basin test setup – Fluke Biomedical ULT800 User Manual
Page 19

Ultrasound Transducer Leakage Current Tester
Using the ULT800
9
ULT800
TEE TRANSDUCER
LEAKAGE CURRENT TESTER
INSTRUCTIONS:
1. Connect the transducer and the adapters to the ULT 800.
2. Press the ON/TEST button and wait for the READY light.
3. Select CONDUCTIVITY and press the ON/TEST Button-
Wait for the PASS light.
4. Select LEAKAGE and press the ON/TEST button again-
observe PASS or FAIL light. If test fails, DO NOT USE
TRANSDUCER. Refer to the user’s manual.
ON
READY
PASS
FAIL
LOW BAT
TEST
CONDUCTIVITY
LEAKAGE
Biomedical
Transducer-Specific
Adapter
Dual Conductivity
Electrodes
Fluke
ULT800
X
ecy02f.eps
Figure 2. Disinfection Basin Test Setup
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)