Ultrasound Transducer Leakage Current Tester

Using the ULT800

9

ULT800

TEE TRANSDUCER
LEAKAGE CURRENT TESTER

INSTRUCTIONS:

1. Connect the transducer and the adapters to the ULT 800.

2. Press the ON/TEST button and wait for the READY light.

3. Select CONDUCTIVITY and press the ON/TEST Button-
Wait for the PASS light.

4. Select LEAKAGE and press the ON/TEST button again-

observe PASS or FAIL light. If test fails, DO NOT USE
TRANSDUCER. Refer to the user’s manual.

ON

READY

PASS

FAIL

LOW BAT

TEST

CONDUCTIVITY

LEAKAGE

Biomedical

Transducer-Specific

Adapter

Dual Conductivity

Electrodes

Fluke
ULT800

X

ecy02f.eps

Figure 2. Disinfection Basin Test Setup